Recall Z-2341-2018

Intuitive Surgical, Inc. · initiated March 7, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

da Vinci X Surgical System, Model IS4200. Product Usage: The Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System Mode IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

Codes / lots: System S/N SL0008, ACJ S/N's P117154256, P117154263, P117154281, P117154286; System S/N SL0012, ACJ S/N's P117154260, P117154297, P117154298, P117154308; System S/N SL0013, ACJ S/N's P117154251, P117154255, P117154257, P117154305; System S/N SL0015, ACJ S/N's P117154274, P117154276, P117154289, P117154290; System S/N SL0016, ACJ S/N's P117154259, P117154291, P117154306, P117154307; System S/N SL0017, ACJ S/N's P117154261, P117154269, P117154277, P117154296; System S/N SL0018, ACJ S/N's P117194009, P117194017, P117194037; System S/N SL0019, ACJ S/N's P117154267, P117154283, P117154294, P117154302; System S/N SL0020, ACJ S/N's P117194008, P117194050, P117194060, P117194078; System S/N SL0021, ACJ S/N's P117194063, P117194064, P117194081, P117194083; System S/N SL0028, ACJ S/N's P117154279, P117154285, P117154292, P117154300; System S/N SL0034, ACJ S/N's P117154264, P117154275, P117154304; System S/N SL0035, ACJ S/N's P117194016, P117194019, P117194070, P117194093; System S/N SL0036, ACJ S/N's P117194025, P117194073, P117241990; System S/N SL0037, ACJ S/N's P117194006, P117194036, P117194056, P117194058 System S/N SL0038, ACJ S/N's P117241971, P117241993; System S/N SL0039, ACJ…

Reason for recall

Specific da Vinci X Surgical Systems (IS4200) were shipped with defective Arm Controller Joint (ACJ) boards that were found to have a high rate of failure due to a defective component.

FDA-determined cause: Nonconforming Material/Component

Action taken

Intuitive Surgical, Inc. sent a Recall letters dated 3/7/2018 to affected customers via FedEx. The letter identified the affected product problem, risk to health, affected serial numbers, the action the firm will be taking, and the actions to be taken by the customer.

Details

Recall number
Z-2341-2018
Initiated
March 7, 2018
Terminated
July 29, 2020
Product code
NAY – System, Surgical, Computer Controlled Instrument
Quantity
108 ACJ's on 35 surgical systems
Distribution
US Nationwide AR, CA, FL, GA, IL, IA, MN, MS, NE, OK, TN, and TX. Foreign distribution was made to Australia, Austria, Finland, France, Germany, Italy, Netherlands, Spain, and United Kingdom.
Recalling firm
Intuitive Surgical, Inc., Sunnyvale, CA

Official FDA recall record