Recall Z-2399-2020

Leica Microsystems, Inc. · initiated May 20, 2020

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Leica ARveo and M530 OH6 microscope systems.

Codes / lots: M530 OH6 Serial Numbers for Model # 10448680 OH6-016215, OH6-042317, OH6-032016, OH6-009815,OH6-044518 and OH6-030016. Arveo Serial Numbers for Model # 10448999 OHA-006318, OHA-010719, OHA-011919, OHA-000818, OHA-009119, OHA-000718, OHA-014919, OHA-016619, OHA-015919, OHA-007818, OHA-013519, OHA-001218, OHA-012219, OHA-017319, OHA-000418, OHA-017119, OHA-013719, OHA-009919, OHA-017219, OHA-017419, OHA-002218, OHA-000318, OHA-014819, OHA-015219, OHA-003718, OHA-001018, OHA-016019, OHA-012919, OHA-012619, OHA-011719, OHA-012119, OHA-010819, OHA-016119, OHA-009219, OHA-002618, OHA-015719, OHA-012019, OHA-009519, OHA-000618, OHA-008518, OHA-004818, OHA-012719, OHA-010619, OHA-001118, OHA-010419, OHA-016319, OHA-016519, OHA-010219, OHA-012519, OHA-015419, OHA-009018, OHA-016419, OHA-016719, OHA-008918, OHA-007618, OHA-009619, OHA-012319, OHA-010519, OHA-005418.

Reason for recall

Possible unintended interruption of surgical procedures due to equipment shutdown, related to changes in voltage.

FDA-determined cause: Other

Action taken

The firm sent letters to consignees on May 20, 2020 advising them that they can continue to use the system pending this field update. Based on our data analysis, the probability of occurrence for this issue (excessive voltage fluctuation and a subsequent microscope shut off) is remote. However, please ensure that a functional check is performed prior to each surgery as described in section 16.1, directions for the operator of the instrument of the user manual.

Details

Recall number
Z-2399-2020
Initiated
May 20, 2020
Terminated
May 13, 2021
Product code
EPT – Microscope, Surgical
Quantity
Domestic distribution 65 devices. International distribution 12 devices. Total of devices distributed 77.
Distribution
Domestic distribution: AL,AR,AZ, CA, CT, DC, DE, FL, IL, KY, LA, MI, MN, MO, NC, NY, OH. OR. TN, TX, UT, VA, WA, WI ,WV. International distribution: Japan
Recalling firm
Leica Microsystems, Inc., Buffalo Grove, IL

Official FDA recall record