Recall Z-2452-2018

Ethicon, Inc. · initiated May 7, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

PROLENE SUTURE 30"(75CM) 4-0 BLUE, D7748

Codes / lots: KDH853 KKH769 KLQ193 KPH698 LCH183 LCH570 LGB761 LJB020 LJH536 MCJ369 MCJ447

Reason for recall

The needle associated with certain suture product codes has the potential to perforate the packaging due to a packaging design issue. A perforated package has the potential to cause a soft tissue injury to a healthcare professional handling the subject product or a patient infection due to a breach of sterility within the packaging if it goes unnoticed by the healthcare professional.

FDA-determined cause: Package design/selection

Action taken

On May 24, 2018, the firm distributed Urgent: Medical Device Recall letters to affected U.S. customers. Customers were instructed to do the following: 1. Examine your inventory immediately to determine if you have product subject to this recall (removal) on hand and quarantine such product(s). 2. Remove the product subject to this recall (removal) and communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. 3. If any product subject to this recall (removal) has been forwarded to another facility, contact that facility to arrange return. 4. Complete the Business Reply Form (BRF) confirming receipt of this notice and fax or email it to Stericycle at 1-888-473-8015 or Ethicon3681@stericycle.com within three (3) business days. Please return the BRF even if you do not have product subject to this recall (removal). 5. Keep this notice visibly posted for awareness until all product subject to this recall (removal) has been returned to Stericycle. While processing your returns, please maintain a copy of this notice with the product subject to this recall (removal) and keep a copy for your records. 6. Customers are required to return all unused D-Special PROLENE Polypropylene Suture & D-Special PRONOVA Poly (Hexafluoropropylene - VDF) Suture products that are in their inventory immediately. Only D-Special PROLENE Polypropylene Suture & D-Special PRONOVA Poly (Hexafluoropropylene - VDF) Suture product subject to this recall (removal) returned by September 21, 2018 will be eligible for credit reimbursement. Any product subject to this recall (removal) returned after September 21, 2018 will not be eligible for credit reimbursement. 7. To return product subject to this recall (removal), photocopy the completed BRF, place it in the box with the product, and affix the pre-paid authorized shipping label included with the recall notification letter. Ethicon will pay for

Details

Recall number
Z-2452-2018
Initiated
May 7, 2018
Terminated
June 8, 2020
Product code
GAW – Suture, Nonabsorbable, Synthetic, Polypropylene
Quantity
2844
Distribution
Accounts in AZ, CA, CO, FL, GA, IL, IN, KS, MA, MN, MO, NC, NE, NJ, NY, OH, OR, SC, SD, TN, TX, VA, WA, WI, WV.
Recalling firm
Ethicon, Inc., Somerville, NJ

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2448-2018 Class IIPROLENE SUTURE MONO BLUE - BV1755 Taper Point 3/8C, D10031The needle associated with certain suture product codes has the potential to perforate the packaging due to… 05/07/2018Terminated
Z-2449-2018 Class IIPROLENE SUTURE BLUE 4X90CM M1.5 USP 4/0 DA RB-1 +P, D5981The needle associated with certain suture product codes has the potential to perforate the packaging due to… 05/07/2018Terminated
Z-2450-2018 Class IIPROLENE SUTURE 4-36"(90CM) 4-0 BLUE, D5988The needle associated with certain suture product codes has the potential to perforate the packaging due to… 05/07/2018Terminated
Z-2451-2018 Class IIPROLENE SUTURE 4-36" (90CM) 3-0 BLUE, D6416The needle associated with certain suture product codes has the potential to perforate the packaging due to… 05/07/2018Terminated
Z-2453-2018 Class IIPROLENE SUTURE 36"(90CM) 4-0 BLUE, D7768The needle associated with certain suture product codes has the potential to perforate the packaging due to… 05/07/2018Terminated
Z-2454-2018 Class IIPROLENE SUTURE BLUE 4X18"5/0 DA RB-2 +PLG, D7879The needle associated with certain suture product codes has the potential to perforate the packaging due to… 05/07/2018Terminated
Z-2455-2018 Class IIPROLENE SUTURE 30"(75CM) 8-0 BLUE, D8292The needle associated with certain suture product codes has the potential to perforate the packaging due to… 05/07/2018Terminated
Z-2456-2018 Class IIPROLENE SUTURE 30"(75CM) 6-0 BLUE, D8647The needle associated with certain suture product codes has the potential to perforate the packaging due to… 05/07/2018Terminated
Z-2457-2018 Class IIPROLENE SUTURE 24"(60CM) 6-0 BLUE, D8688The needle associated with certain suture product codes has the potential to perforate the packaging due to… 05/07/2018Terminated
Z-2458-2018 Class IIPROLENE SUTURE 24"(60CM) 7-0 BLUE, D8881The needle associated with certain suture product codes has the potential to perforate the packaging due to… 05/07/2018Terminated
Z-2459-2018 Class IIPROLENE SUTURE 36"(90CM) 4-0 BLUE, D9067The needle associated with certain suture product codes has the potential to perforate the packaging due to… 05/07/2018Terminated
Z-2460-2018 Class IIPROLENE SUTURE 4"(10CM) 8-0 BLUE, D9662The needle associated with certain suture product codes has the potential to perforate the packaging due to… 05/07/2018Terminated
Z-2461-2018 Class IIPROLENE SUTURE BLUE 48" 4-0 D/A SH-1, D9746The needle associated with certain suture product codes has the potential to perforate the packaging due to… 05/07/2018Terminated
Z-2462-2018 Class IIPROLENE SUTURE 36"(90CM) 3-0 BLU, D9844The needle associated with certain suture product codes has the potential to perforate the packaging due to… 05/07/2018Terminated
Z-2463-2018 Class IIPROLENE SUTURE MONO BLUE BV1755 Taperpoint 3/8C, D9995The needle associated with certain suture product codes has the potential to perforate the packaging due to… 05/07/2018Terminated
Z-2464-2018 Class IIPROLENE SUTURE BLUE 18" 7-0 D/A BV-1, DC631The needle associated with certain suture product codes has the potential to perforate the packaging due to… 05/07/2018Terminated
Z-2465-2018 Class IIPRONOVA SUTURE BLUE MONO POLY - SH-1 Taperpoint 1/2C, D10027The needle associated with certain suture product codes has the potential to perforate the packaging due to… 05/07/2018Terminated
Z-2466-2018 Class IIPRONOVA SUTURE BLUE 122CM M1.5, D10189The needle associated with certain suture product codes has the potential to perforate the packaging due to… 05/07/2018Terminated