Recall Z-2454-2018
Ethicon, Inc. · initiated May 7, 2018
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
PROLENE SUTURE BLUE 4X18"5/0 DA RB-2 +PLG, D7879
Codes / lots: KEB378 KKJ784 MCP374
Reason for recall
The needle associated with certain suture product codes has the potential to perforate the packaging due to a packaging design issue. A perforated package has the potential to cause a soft tissue injury to a healthcare professional handling the subject product or a patient infection due to a breach of sterility within the packaging if it goes unnoticed by the healthcare professional.
FDA-determined cause: Package design/selection
Action taken
On May 24, 2018, the firm distributed Urgent: Medical Device Recall letters to affected U.S. customers. Customers were instructed to do the following: 1. Examine your inventory immediately to determine if you have product subject to this recall (removal) on hand and quarantine such product(s). 2. Remove the product subject to this recall (removal) and communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. 3. If any product subject to this recall (removal) has been forwarded to another facility, contact that facility to arrange return. 4. Complete the Business Reply Form (BRF) confirming receipt of this notice and fax or email it to Stericycle at 1-888-473-8015 or Ethicon3681@stericycle.com within three (3) business days. Please return the BRF even if you do not have product subject to this recall (removal). 5. Keep this notice visibly posted for awareness until all product subject to this recall (removal) has been returned to Stericycle. While processing your returns, please maintain a copy of this notice with the product subject to this recall (removal) and keep a copy for your records. 6. Customers are required to return all unused D-Special PROLENE Polypropylene Suture & D-Special PRONOVA Poly (Hexafluoropropylene - VDF) Suture products that are in their inventory immediately. Only D-Special PROLENE Polypropylene Suture & D-Special PRONOVA Poly (Hexafluoropropylene - VDF) Suture product subject to this recall (removal) returned by September 21, 2018 will be eligible for credit reimbursement. Any product subject to this recall (removal) returned after September 21, 2018 will not be eligible for credit reimbursement. 7. To return product subject to this recall (removal), photocopy the completed BRF, place it in the box with the product, and affix the pre-paid authorized shipping label included with the recall notification letter. Ethicon will pay for
Details
- Recall number
- Z-2454-2018
- Initiated
- May 7, 2018
- Terminated
- June 8, 2020
- Product code
- GAW – Suture, Nonabsorbable, Synthetic, Polypropylene
- Quantity
- 588
- Distribution
- Accounts in AZ, CA, CO, FL, GA, IL, IN, KS, MA, MN, MO, NC, NE, NJ, NY, OH, OR, SC, SD, TN, TX, VA, WA, WI, WV.
- Recalling firm
- Ethicon, Inc., Somerville, NJ
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2448-2018 Class II | PROLENE SUTURE MONO BLUE - BV1755 Taper Point 3/8C, D10031The needle associated with certain suture product codes has the potential to perforate the packaging due to… | 05/07/2018Terminated |
| Z-2449-2018 Class II | PROLENE SUTURE BLUE 4X90CM M1.5 USP 4/0 DA RB-1 +P, D5981The needle associated with certain suture product codes has the potential to perforate the packaging due to… | 05/07/2018Terminated |
| Z-2450-2018 Class II | PROLENE SUTURE 4-36"(90CM) 4-0 BLUE, D5988The needle associated with certain suture product codes has the potential to perforate the packaging due to… | 05/07/2018Terminated |
| Z-2451-2018 Class II | PROLENE SUTURE 4-36" (90CM) 3-0 BLUE, D6416The needle associated with certain suture product codes has the potential to perforate the packaging due to… | 05/07/2018Terminated |
| Z-2452-2018 Class II | PROLENE SUTURE 30"(75CM) 4-0 BLUE, D7748The needle associated with certain suture product codes has the potential to perforate the packaging due to… | 05/07/2018Terminated |
| Z-2453-2018 Class II | PROLENE SUTURE 36"(90CM) 4-0 BLUE, D7768The needle associated with certain suture product codes has the potential to perforate the packaging due to… | 05/07/2018Terminated |
| Z-2455-2018 Class II | PROLENE SUTURE 30"(75CM) 8-0 BLUE, D8292The needle associated with certain suture product codes has the potential to perforate the packaging due to… | 05/07/2018Terminated |
| Z-2456-2018 Class II | PROLENE SUTURE 30"(75CM) 6-0 BLUE, D8647The needle associated with certain suture product codes has the potential to perforate the packaging due to… | 05/07/2018Terminated |
| Z-2457-2018 Class II | PROLENE SUTURE 24"(60CM) 6-0 BLUE, D8688The needle associated with certain suture product codes has the potential to perforate the packaging due to… | 05/07/2018Terminated |
| Z-2458-2018 Class II | PROLENE SUTURE 24"(60CM) 7-0 BLUE, D8881The needle associated with certain suture product codes has the potential to perforate the packaging due to… | 05/07/2018Terminated |
| Z-2459-2018 Class II | PROLENE SUTURE 36"(90CM) 4-0 BLUE, D9067The needle associated with certain suture product codes has the potential to perforate the packaging due to… | 05/07/2018Terminated |
| Z-2460-2018 Class II | PROLENE SUTURE 4"(10CM) 8-0 BLUE, D9662The needle associated with certain suture product codes has the potential to perforate the packaging due to… | 05/07/2018Terminated |
| Z-2461-2018 Class II | PROLENE SUTURE BLUE 48" 4-0 D/A SH-1, D9746The needle associated with certain suture product codes has the potential to perforate the packaging due to… | 05/07/2018Terminated |
| Z-2462-2018 Class II | PROLENE SUTURE 36"(90CM) 3-0 BLU, D9844The needle associated with certain suture product codes has the potential to perforate the packaging due to… | 05/07/2018Terminated |
| Z-2463-2018 Class II | PROLENE SUTURE MONO BLUE BV1755 Taperpoint 3/8C, D9995The needle associated with certain suture product codes has the potential to perforate the packaging due to… | 05/07/2018Terminated |
| Z-2464-2018 Class II | PROLENE SUTURE BLUE 18" 7-0 D/A BV-1, DC631The needle associated with certain suture product codes has the potential to perforate the packaging due to… | 05/07/2018Terminated |
| Z-2465-2018 Class II | PRONOVA SUTURE BLUE MONO POLY - SH-1 Taperpoint 1/2C, D10027The needle associated with certain suture product codes has the potential to perforate the packaging due to… | 05/07/2018Terminated |
| Z-2466-2018 Class II | PRONOVA SUTURE BLUE 122CM M1.5, D10189The needle associated with certain suture product codes has the potential to perforate the packaging due to… | 05/07/2018Terminated |