Recall Z-2661-2020
Becton Dickinson & Company · initiated June 19, 2020
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
PowerLoc MAX Power-Injectable Infusion Set myPICK Port Access Kit; Catalog Numbers CPA00001 (UDI 00801741098345), CPA00002 (UDI 00801741098352), CPA00028 (UDI 00801741129728), CPA00032 (UDI 00801741137501), CPA00033 (UDI 00801741137501), CPA00036 (UDI 00801741138911), CPA00051 (UDI 00801741175718), CPA00055 UDI 00801741182181), CPA00056 (UDI 00801741182907)
Codes / lots: Catalog Numbers CPA00001 (UDI 00801741098345), CPA00002 (UDI 00801741098352), CPA00028 (UDI 00801741129728), CPA00032 (UDI 00801741137501), CPA00033 (UDI 00801741137501), CPA00036 (UDI 00801741138911), CPA00051 (UDI 00801741175718), CPA00055 UDI 00801741182181), CPA00056 (UDI 00801741182907). Lot Numbers: REEN0212 REEP2685 REEP3660 REEQ3563 REER1968 REDZ0705 REEP3313 REEP3547 REER0577 REEN0214 REEQ3569 REDY3552 REEN1934 REEP3280 REEP3552 REER1115 REEN0218 REEP3659 REEP1436 REER0298 REER1335 REDY0992 REEN0278 REEP1437 REEQ0232
Reason for recall
Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.
FDA-determined cause: Process control
Action taken
URGENT MEDICAL DEVICE RECALL notification letters dated 6/19/20 were sent to customers. PLEASE TAKE THE FOLLOWING ACTIONS: 1. Immediately review your inventory for the specific catalog and lot numbers referenced in Attachment A. Discontinue use of those devices and destroy all product subject to this recall following your institutions process for destruction. MDS-20-1979 Page 2 of 28 BD Medical 1 Becton Drive Franklin Lakes, NJ 07417 USA bd.com 2. Share this notification with all users of the product within your facility network to ensure they are also aware of this recall. 3. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements and subsequently process your credit accordingly for the discarded material. a. Customers who purchased product directly from BD will receive credit. b. Customers who did not purchase product directly from BD will need to contact their distributor to obtain credit. 4. Report any adverse health consequences experienced with the use of this product to BD at product.complaints@bd.com. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1800FDA1088 (18003321088) Mail: MedWatch, HF2, FDA, 5600 Fishers Lane, Rockville, MD 208529787 Customer/Technical Support Phone: 8002901689 Monday Friday between 8:00 am to 5:00 pm MDT SLCBASD.FieldAction@bd.com An Urgent Medical Device Recall notification letter dated 8/25/20 was sent to customers informing them of two additional lots of affected product. An URGENT MEDICAL DEVICE RECALL UPDATE dated 10/7/20 was sent to customers informing them of two additional lots of affected product.
Details
- Recall number
- Z-2661-2020
- Initiated
- June 19, 2020
- Posted by FDA
- July 20, 2020
- Terminated
- July 29, 2022
- Product code
- PTI – Non-Coring (Huber) Needle
- Quantity
- 5,296,132 total
- Distribution
- Worldwide Distribution
- Recalling firm
- Becton Dickinson & Company, Franklin Lakes, NJ
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2660-2020 Class II | PowerLoc MAX Power Injectable Infusion Set AllPoints Port Access System; Catalog Numbers…Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that… | 06/19/2020Terminated |
| Z-2662-2020 Class II | PowerLoc MAX Power Injectable Infusion Port Access Kit; Catalog Numbers 2131910 (UDI…Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that… | 06/19/2020Terminated |
| Z-2663-2020 Class II | PowerLoc MAX Power Injectable Infusion Set with GuardIVa Antimicrobial Hemostatic…Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that… | 06/19/2020Terminated |
| Z-2664-2020 Class II | SafeStep Huber Needle Set AllPoints Port Access System; Catalog Numbers AN152075 (UDI…Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that… | 06/19/2020Terminated |
| Z-2665-2020 Class II | SafeStep Huber Needle Set myPICK Port Access Kit; Catalog Numbers CPA00020A (UDI…Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that… | 06/19/2020Terminated |
| Z-2666-2020 Class II | SafeStep Huber Needle Set Port Access Kit; Catalog Numbers PA-0029 (UDI 00801741085482)…Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that… | 06/19/2020Terminated |
| Z-2667-2020 Class II | SafeStep Huber Needle Set with GuardIVa Antimicrobial Hemostatic Dressing; Catalog…Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that… | 06/19/2020Terminated |
| Z-2668-2020 Class II | SafeStep Huber Needle Set with GuardIVa Antimicrobial Hemostatic Dressing Port Access…Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that… | 06/19/2020Terminated |
| Z-2669-2020 Class II | SafeStep Huber Needle Set with Y Injection Site and Medegen Needleless Injection Cap…Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that… | 06/19/2020Terminated |
| Z-2670-2020 Class II | PowerLoc MAX Power Injectable Infusion Set; Catalog Numbers 0131910 (UDI…Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that… | 06/19/2020Terminated |
| Z-2671-2020 Class II | SafeStep Huber Needle Set; Catalog Numbers LH-0029 (UDI 00801741066160), LH-0030 (UDI…Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that… | 06/19/2020Terminated |