Recall Z-2670-2020

Becton Dickinson & Company · initiated June 19, 2020

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

PowerLoc MAX Power Injectable Infusion Set; Catalog Numbers 0131910 (UDI 00801741047329), 0131915 (UDI 00801741047336), 0131975 (UDI 00801741047343), 0132010 (UDI 00801741047350), 0132015 (UDI 00801741047367), 0132075 (UDI 00801741047374), 0132210 (UDI 00801741047381), 0132215 (UDI 00801741047398), 0132275 (UDI 00801741047404), 0141910 (UDI 00801741047411), 0141915 (UDI 00801741047428), 0141975 (UDI 00801741047435), 0142010 (UDI 00801741047442), 0142015 (UDI 00801741047459), 0142075 (UDI 00801741047466), 0142210 (UDI 00801741047473), 0142215 (UDI 00801741047480), 0142275 (UDI 00801741047497)

Codes / lots: Catalog Numbers 0131910 (UDI 00801741047329), 0131915 (UDI 00801741047336), 0131975 (UDI 00801741047343), 0132010 (UDI 00801741047350), 0132015 (UDI 00801741047367), 0132075 (UDI 00801741047374), 0132210 (UDI 00801741047381), 0132215 (UDI 00801741047398), 0132275 (UDI 00801741047404), 0141910 (UDI 00801741047411), 0141915 (UDI 00801741047428), 0141975 (UDI 00801741047435), 0142010 (UDI 00801741047442), 0142015 (UDI 00801741047459), 0142075 (UDI 00801741047466), 0142210 (UDI 00801741047473), 0142215 (UDI 00801741047480), 0142275 (UDI 00801741047497). Lot Numbers: ASCXF007 ASCXF019 ASDRF074 ASDRF075 ASDRF079 ASDRF084 ASDVF021 ASCXF001 ASDNF040 ASDNF043 ASDVF031 ASDWF118 ASCXF006 ASCXF018 ASDRF053 ASDRF058 ASDRF060 ASDRF071 ASDVF029 ASDWF117 ASCYF028 ASCYF029 ASCZF025 ASDPF029 ASDQF015 ASDQF039 ASDQF041 ASDRF065 ASDTF043 ASDTF054 ASDTF055 ASDTF061 ASDVF018 ASCYF023 ASCYF024 ASDTF031 ASDTF032 ASDTF033 ASDTF037 ASDTF038 ASCYF040 ASCYF076 ASCZF055 ASDNF001 ASDNF002 ASDNF004 ASDNF018 ASDQF036 ASDTF075 ASDTF096 ASDTF104 ASDTF105 ASDTF115 ASDUF006 ASDUF019 ASDUF026 ASDUF033 ASDUF042 ASDUF059 ASDUF063 ASDUF078 ASDUF097 ASDUF102 ASDUF105 ASDUF111 ASDUF119 ASDVF005 ASDVF014 ASDWF070 ASDWF081…

Reason for recall

Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.

FDA-determined cause: Process control

Action taken

URGENT MEDICAL DEVICE RECALL notification letters dated 6/19/20 were sent to customers. PLEASE TAKE THE FOLLOWING ACTIONS: 1. Immediately review your inventory for the specific catalog and lot numbers referenced in Attachment A. Discontinue use of those devices and destroy all product subject to this recall following your institutions process for destruction. MDS-20-1979 Page 2 of 28 BD Medical 1 Becton Drive Franklin Lakes, NJ 07417 USA bd.com 2. Share this notification with all users of the product within your facility network to ensure they are also aware of this recall. 3. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements and subsequently process your credit accordingly for the discarded material. a. Customers who purchased product directly from BD will receive credit. b. Customers who did not purchase product directly from BD will need to contact their distributor to obtain credit. 4. Report any adverse health consequences experienced with the use of this product to BD at product.complaints@bd.com. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1800FDA1088 (18003321088) Mail: MedWatch, HF2, FDA, 5600 Fishers Lane, Rockville, MD 208529787 Customer/Technical Support Phone: 8002901689 Monday Friday between 8:00 am to 5:00 pm MDT SLCBASD.FieldAction@bd.com An Urgent Medical Device Recall notification letter dated 8/25/20 was sent to customers informing them of two additional lots of affected product. An URGENT MEDICAL DEVICE RECALL UPDATE dated 10/7/20 was sent to customers informing them of two additional lots of affected product.

Details

Recall number
Z-2670-2020
Initiated
June 19, 2020
Posted by FDA
July 20, 2020
Terminated
July 29, 2022
Product code
PTI – Non-Coring (Huber) Needle
Quantity
5,296,132 total
Distribution
Worldwide Distribution
Recalling firm
Becton Dickinson & Company, Franklin Lakes, NJ

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2660-2020 Class IIPowerLoc MAX Power Injectable Infusion Set AllPoints Port Access System; Catalog Numbers…Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that… 06/19/2020Terminated
Z-2661-2020 Class IIPowerLoc MAX Power-Injectable Infusion Set myPICK Port Access Kit; Catalog Numbers…Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that… 06/19/2020Terminated
Z-2662-2020 Class IIPowerLoc MAX Power Injectable Infusion Port Access Kit; Catalog Numbers 2131910 (UDI…Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that… 06/19/2020Terminated
Z-2663-2020 Class IIPowerLoc MAX Power Injectable Infusion Set with GuardIVa Antimicrobial Hemostatic…Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that… 06/19/2020Terminated
Z-2664-2020 Class IISafeStep Huber Needle Set AllPoints Port Access System; Catalog Numbers AN152075 (UDI…Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that… 06/19/2020Terminated
Z-2665-2020 Class IISafeStep Huber Needle Set myPICK Port Access Kit; Catalog Numbers CPA00020A (UDI…Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that… 06/19/2020Terminated
Z-2666-2020 Class IISafeStep Huber Needle Set Port Access Kit; Catalog Numbers PA-0029 (UDI 00801741085482)…Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that… 06/19/2020Terminated
Z-2667-2020 Class IISafeStep Huber Needle Set with GuardIVa Antimicrobial Hemostatic Dressing; Catalog…Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that… 06/19/2020Terminated
Z-2668-2020 Class IISafeStep Huber Needle Set with GuardIVa Antimicrobial Hemostatic Dressing Port Access…Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that… 06/19/2020Terminated
Z-2669-2020 Class IISafeStep Huber Needle Set with Y Injection Site and Medegen Needleless Injection Cap…Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that… 06/19/2020Terminated
Z-2671-2020 Class IISafeStep Huber Needle Set; Catalog Numbers LH-0029 (UDI 00801741066160), LH-0030 (UDI…Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that… 06/19/2020Terminated