Recall Z-2796-2026

AVID Medical, Inc. · initiated April 30, 2026

Class I Open, Classified

Class I – dangerous or defective products that could cause serious health problems or death

Product

Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. CARDIAC CATH PACK EAMC1000-05

Codes / lots: UDI-DI 10809160377314 Lots 1573593 1583827 1609541 1612134 1612516 1619132 1623384 1627013 1635459 1672306 1675159

Reason for recall

Affected kits contain Medline Namic Angiographic Star Off Handle Manifolds, which were recalled by the manufacturer due to presence of particulates within fluid path of the Manifolds. Particulate within the fluid path has the potential to be introduced into blood circulation and become lodged within blood vessels.

FDA-determined cause: Under Investigation by firm

Action taken

On April 30, 2026, firm notified distributors via email. Distributors were tasked with notifying their end-users. Customers were instructed to identify, segregate, and quarantine all impacted products. They were provided with a labeling template to print warning labels which should be affixed on all affected kits and cases in inventory. The firm is also able to provide pre-printed labels if needed. Warning labels inform users to remove and discard the manifold prior to use.

Details

Recall number
Z-2796-2026
Initiated
April 30, 2026
Posted by FDA
July 31, 2026
Product code
OES – Cardiac Catheterization Kit
Quantity
5168 kits
Distribution
US Nationwide distribution in the states of NC and TX.
Recalling firm
AVID Medical, Inc., Toano, VA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2797-2026 Class IHalyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle…Affected kits contain Medline Namic Angiographic Star Off Handle Manifolds, which were recalled by the… 04/30/2026Open, Classified
Z-2798-2026 Class IHalyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle…Affected kits contain Medline Namic Angiographic Star Off Handle Manifolds, which were recalled by the… 04/30/2026Open, Classified