AVID Medical, Inc. · initiated April 30, 2026

Class I Open, Classified

Class I – dangerous or defective products that could cause serious health problems or death

Product

Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. HEART CATH, SELF REGIONAL SELF131-05

Codes / lots: UDI-DI 10809160408650 Lots 1618351 1621441 1627930 1635908 1644501 1646021 1651957 1654887 1656581 1656582 1659713 1670884 1670885 1671950

Reason for recall

Affected kits contain Medline Namic Angiographic Star Off Handle Manifolds, which were recalled by the manufacturer due to presence of particulates within fluid path of the Manifolds. Particulate within the fluid path has the potential to be introduced into blood circulation and become lodged within blood vessels.

FDA-determined cause: Under Investigation by firm

Action taken

On April 30, 2026, firm notified distributors via email. Distributors were tasked with notifying their end-users. Customers were instructed to identify, segregate, and quarantine all impacted products. They were provided with a labeling template to print warning labels which should be affixed on all affected kits and cases in inventory. The firm is also able to provide pre-printed labels if needed. Warning labels inform users to remove and discard the manifold prior to use.

Details

Recall number
Z-2798-2026
Initiated
April 30, 2026
Posted by FDA
July 31, 2026
Product code
OEZ – Cardiovascular Procedure Kit
Quantity
5168 kits
Distribution
US Nationwide distribution in the states of NC and TX.
Recalling firm
AVID Medical, Inc., Toano, VA