Recall Z-2821-2026
Instrumentation Laboratory · initiated May 8, 2026
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Intelligent Quality Management 2 (iQM2) is used as the quality control and assessment system for the GEM Premier 5000 system.
Codes / lots: Lot Code: GEM Premier 5000: Part number: 00024019255 UDI: 08426950870227 Units distributed in US: 0 affected BCR wands on the GEM Premier 5000 in the US. GEM Premier 5000 Serial Nos.: N/A Units distributed in Hong Kong: 27 affected BCR wands on the GEM Premier 5000 in Hong Kong. GEM Premier 5000 Serial Nos.: 20106073, 20106039, 19052933, 20116415, 20064792, 22062065, 22062067, 19103490, 22092716, 22102881, 18112339, 19012518, 20126582, 20075256, 20064794, 20106036, 24016057, 21088758, 21088759, 18061910, 24016081, 24016108, 24108036, 24108044, 24067370, 24067371 and 24077468. Units distributed in United Kingdom: 67 affected BCR wands on the GEM Premier 5000 in the United Kingdom. GEM Premier 5000 Serial Nos.: 25110547, 25100462, 25110522, 25100472, 25101897, 25101920, 25101931, 25122029, 26012032, 25122030, 26012033, 26012043, 25110523, 25110536, 25120550, 25069638, 23101109, 24011137, 25081799, 25081797, 25091839, 25091840, 25091864, 25091870, 25091868, 25091873, 25091880, 25091876, 25101886, 25101889, 25101887, 25101891, 25101907, 25101904, 25101912, 25101905, 25101917, 25101921, 25101926, 25101929, 25101934, 25101927, 25111968, 25121986, 25121985, 25121981, 25121984, 25121994…
Reason for recall
BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after April 8, 2026) may assign patient results to an incorrect patient ID. The root cause is a race condition introduced when the "price check" feature was disabled to reduce excess current draw. This causes buffered data from a previous scan to be transmitted instead of the most recent scan, populating a prior patient ID instead of the current one. This can result in erroneous results being attributed to the wrong patient, potentially leading to inappropriate clinical management.
FDA-determined cause: Software Design Change
Action taken
Werfen dba Instrumentation laboratory notified consignees on about 05/08/2026 via letter and/or email. Consignees were instructed to discontinue use of the affected BCR wands immediately, manually enter patient IDs when performing testing on the GEM Premier 5000 and/or 7000 until the BCR wands have been replaced, contact Werfen Service at 1-800-955-9525 to arrange for prompt replacement of the affected BCR wands, post this notification on or near affected GEM Premier 5000 and/or 7000 instruments, and inform all laboratory staff of this notification. Consignees were also instructed to forward the notification to all affected locations within each facility, retain a copy of the notification for records, and complete and return the Mandatory Response Tracking Form within 10 days of receipt. In addition to the tracked customer notification, a Service Information (SI) 2026-09 was issued on May 11, 2026, for service personnel to track the destruction of the affected BCR wands. The Service Information includes a destruction form for service personnel to document the destruction of the affected BCR wands and to return the form to Regulatory for tracking purposes.
Details
- Recall number
- Z-2821-2026
- Initiated
- May 8, 2026
- Posted by FDA
- July 29, 2026
- Product code
- CHL – Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
- Quantity
- 0 US, 118 OUS
- Distribution
- US Nationwide distribution.
- Recalling firm
- Instrumentation Laboratory, Bedford, MA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2822-2026 Class II | Brand Name: GEM Premier 7000 with iQM3 Product Name: GEM Premier 7000 with iQM3…BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after… | 05/08/2026Open, Classified |
| Z-2823-2026 Class II | Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number:…BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after… | 05/08/2026Open, Classified |
| Z-2824-2026 Class II | Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number:…BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after… | 05/08/2026Open, Classified |
| Z-2825-2026 Class II | Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number:…BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after… | 05/08/2026Open, Classified |
| Z-2826-2026 Class II | Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number:…BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after… | 05/08/2026Open, Classified |