Recall Z-2822-2026

Instrumentation Laboratory · initiated May 8, 2026

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Brand Name: GEM Premier 7000 with iQM3 Product Name: GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Intelligent Quality Management 3 (iQM3) is used as the quality control (QC) and assessment system for the GEM Premier 7000.

Codes / lots: Lot Code: GEM Premier 7000: Part number: 00000015279 UDI: 0195226000924 Units distributed in US: 4 affected BCR wands on the GEM Premier 7000 in the US. GEM Premier 7000 Serial Nos. 26012062, 26012065, 26012069 and 26012070. Affected Accessory: Bar Code Reader (BCR) Wand Part number: 00024015859 UDI: 08426950869634 Units distributed in US: 4

Reason for recall

BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after April 8, 2026) may assign patient results to an incorrect patient ID. The root cause is a race condition introduced when the "price check" feature was disabled to reduce excess current draw. This causes buffered data from a previous scan to be transmitted instead of the most recent scan, populating a prior patient ID instead of the current one. This can result in erroneous results being attributed to the wrong patient, potentially leading to inappropriate clinical management.

FDA-determined cause: Software Design Change

Action taken

Werfen dba Instrumentation laboratory notified consignees on about 05/08/2026 via letter and/or email. Consignees were instructed to discontinue use of the affected BCR wands immediately, manually enter patient IDs when performing testing on the GEM Premier 5000 and/or 7000 until the BCR wands have been replaced, contact Werfen Service at 1-800-955-9525 to arrange for prompt replacement of the affected BCR wands, post this notification on or near affected GEM Premier 5000 and/or 7000 instruments, and inform all laboratory staff of this notification. Consignees were also instructed to forward the notification to all affected locations within each facility, retain a copy of the notification for records, and complete and return the Mandatory Response Tracking Form within 10 days of receipt. In addition to the tracked customer notification, a Service Information (SI) 2026-09 was issued on May 11, 2026, for service personnel to track the destruction of the affected BCR wands. The Service Information includes a destruction form for service personnel to document the destruction of the affected BCR wands and to return the form to Regulatory for tracking purposes.

Details

Recall number
Z-2822-2026
Initiated
May 8, 2026
Posted by FDA
July 29, 2026
Product code
CHL – Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
Quantity
4 US
Distribution
US Nationwide distribution.
Recalling firm
Instrumentation Laboratory, Bedford, MA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2821-2026 Class IIBrand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number:…BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after… 05/08/2026Open, Classified
Z-2823-2026 Class IIBrand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number:…BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after… 05/08/2026Open, Classified
Z-2824-2026 Class IIBrand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number:…BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after… 05/08/2026Open, Classified
Z-2825-2026 Class IIBrand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number:…BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after… 05/08/2026Open, Classified
Z-2826-2026 Class IIBrand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number:…BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after… 05/08/2026Open, Classified