Recall Z-2958-2024

Abiomed, Inc. · initiated August 5, 2024

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Impella CP Smart Assist Set, Canada; Product Code: 0048-0044;

Codes / lots: Product Code: 0048-0044; UDI-DI: 00813502012200; Serial Numbers: 383487 383488 383489 381637 381777 381640 381639 381638 379660 379659 379661 372534 372535 386882 389812 389809 389811 389814 389813 386881 386879 386880 386943 388495 388498 391472 391470 391471 387330 387329 387333 387332 387331 389489 389493 389491 389490 389492 387328 388467 388465 388527 388528 388469 388470 388471 386005 386001 386008 386006 386007 386004 386170 386168 386169 386003 385999 386009 385933 385931 385921 386000 385938 385939 385940 385923 385922 386010 385935 385937 385934 390915 390914 390913 390912 392642 392643 392644 392534 392535 392537 392538 392640 392641 392536 392835 392834 392829 392828 392903 392902 392836 392837 392833 392906 392824 392831 392830 392826 392825 393083 393082 393079 393080 393081 427384 427383 427382 427381 427379 428638 428639 428640 428630 428641 428642 428631 428629 428623 428632 429793 429780 429794 429784 429786 429787 429788 430901 429789 428645 428646 429770 431699 431702 431700 431701 431695 430897 430898 430899 433880 429771 433881 431694 431696 431698 430900 430904 430903 430902 429792 429785 429782 429797 429791 429790 435973 435975 435974 435976 435149 435152…

Reason for recall

A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.

FDA-determined cause: Under Investigation by firm

Action taken

On August 5, 2024 URGENT MEDICAL DEVICE CORRECTION (NOTIFICATION) letters were sent to consignees. ACTIONS TO BE TAKEN BY CUSTOMER/USER: " Product is NOT being removed from the field and does not need to be returned. " Review, complete all fields, sign, and return the attached business response form (BRF) (refer to Attachment 2) to Abiomed2860@SEDGWICK.COM. " Forward this notice to anyone in your facility that needs to be informed (i.e., those who manage, transport, store, stock, or use the subject products). " If any of the subject products have been forwarded to another facility, contact that facility and provide them with this notice. " Post a copy of this notice in a visible area for awareness of this field safety notice. " As with any medical device, adverse reactions or quality problems experienced with the use of this product should be reported to the FDA s MedWatch Adverse Event Reporting Program using the link below: https://www.fda.gov/safety/medwatch-fda-safety-information-and-adverse-event-reporting-program/reporting-serious-problems-fda At Abiomed, our priority is to our customers and their patients, and that includes the safe and effective use of our products. If you have questions or concerns regarding this notice, please contact your local clinical field staff. Thank you for your cooperation.

Details

Recall number
Z-2958-2024
Initiated
August 5, 2024
Posted by FDA
September 4, 2024
Product code
OZD – Temporary Non-Roller Type Left Heart Support Blood Pump
Quantity
323 units
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Brunei Darussalam, Canada, Croatia, Czeck Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Panama, Poland, Saudi Arabia, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, U.A.E.…
Recalling firm
Abiomed, Inc., Danvers, MA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2954-2024 Class IIImpella CP Smart Assist Set, EU; Product Code: 0048-0014;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2955-2024 Class IIImpella CP Smart Assist Set, JP; Product Code: 0048-0024-JP;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2956-2024 Class IIImpella CP Pump Set, Japan; Product Code: 0048-0034-JP;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2957-2024 Class IIImpella CP Pump 371 Set, Clinical Trial; Product Code: 0048-0040;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2959-2024 Class IIPump 371 14F LT CMR Set; Product Code: 0048-0045;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2960-2024 Class IIImpella CP Smart Assist Set APAC; Product Code: 0048-0047;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2961-2024 Class IIImpella 5.5 with SmartAssist Set, EU; Product Code: 0550-0002;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2962-2024 Class IIImpella 5.5 with SmartAssist Set, CA; Product Code: 0550-0004;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2963-2024 Class IIImpella 5.5 with SmartAssist Set, US; Product Code: 0550-0008;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2964-2024 Class IIPump 381 Pump Set (US); Product Code: 1000080;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2965-2024 Class IIImpella 5.5 with SmartAssist S2 Set, US; Product Code: 1000100;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2966-2024 Class IIImpella 5.5 SmartAssist Set, JP; Product Code: 1000211;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2967-2024 Class IIImpellaCP SmartAssistSetAPAC, Delo1895; Product Code: 1000302;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2968-2024 Class IIImpella 5.5 Set AU; Product Code: 1000361;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2969-2024 Class IIImpella CP Smart Assist Set; Product Code: 1000402;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2970-2024 Class IIImpella 5.5 SmartAssist S2 Set, JP; Product Code: 1000459;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2971-2024 Class IIImpella 5.5 with SmartAssist S2 Set, EU; Product Code: 1000482;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified