Recall Z-2968-2024

Abiomed, Inc. · initiated August 5, 2024

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Impella 5.5 Set AU; Product Code: 1000361;

Codes / lots: Product Code: 1000361; UDI-DI: 00813502012927; Serial Numbers: 425570 425332 425331 463954 463955 476358 476357 481131 481130 481129 485279 485278 485280 480079 483437 466172A 491829 491830 491831 491832; Batch Numbers: 2024286281 2024286289 2024286292 2024375505 2024375511 2024398832 2024398833 2024413728 2024413879 2024413882 2024431078 2024431158 2024431171 2024431175 2024431650 2024431652 2024431653 2024431656 2024431673 2024431720;

Reason for recall

A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.

FDA-determined cause: Under Investigation by firm

Action taken

On August 5, 2024 URGENT MEDICAL DEVICE CORRECTION (NOTIFICATION) letters were sent to consignees. ACTIONS TO BE TAKEN BY CUSTOMER/USER: " Product is NOT being removed from the field and does not need to be returned. " Review, complete all fields, sign, and return the attached business response form (BRF) (refer to Attachment 2) to Abiomed2860@SEDGWICK.COM. " Forward this notice to anyone in your facility that needs to be informed (i.e., those who manage, transport, store, stock, or use the subject products). " If any of the subject products have been forwarded to another facility, contact that facility and provide them with this notice. " Post a copy of this notice in a visible area for awareness of this field safety notice. " As with any medical device, adverse reactions or quality problems experienced with the use of this product should be reported to the FDA s MedWatch Adverse Event Reporting Program using the link below: https://www.fda.gov/safety/medwatch-fda-safety-information-and-adverse-event-reporting-program/reporting-serious-problems-fda At Abiomed, our priority is to our customers and their patients, and that includes the safe and effective use of our products. If you have questions or concerns regarding this notice, please contact your local clinical field staff. Thank you for your cooperation.

Details

Recall number
Z-2968-2024
Initiated
August 5, 2024
Posted by FDA
September 4, 2024
Product code
OZD – Temporary Non-Roller Type Left Heart Support Blood Pump
Quantity
20 units
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Brunei Darussalam, Canada, Croatia, Czeck Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Panama, Poland, Saudi Arabia, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, U.A.E.…
Recalling firm
Abiomed, Inc., Danvers, MA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2954-2024 Class IIImpella CP Smart Assist Set, EU; Product Code: 0048-0014;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2955-2024 Class IIImpella CP Smart Assist Set, JP; Product Code: 0048-0024-JP;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2956-2024 Class IIImpella CP Pump Set, Japan; Product Code: 0048-0034-JP;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2957-2024 Class IIImpella CP Pump 371 Set, Clinical Trial; Product Code: 0048-0040;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2958-2024 Class IIImpella CP Smart Assist Set, Canada; Product Code: 0048-0044;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2959-2024 Class IIPump 371 14F LT CMR Set; Product Code: 0048-0045;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2960-2024 Class IIImpella CP Smart Assist Set APAC; Product Code: 0048-0047;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2961-2024 Class IIImpella 5.5 with SmartAssist Set, EU; Product Code: 0550-0002;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2962-2024 Class IIImpella 5.5 with SmartAssist Set, CA; Product Code: 0550-0004;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2963-2024 Class IIImpella 5.5 with SmartAssist Set, US; Product Code: 0550-0008;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2964-2024 Class IIPump 381 Pump Set (US); Product Code: 1000080;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2965-2024 Class IIImpella 5.5 with SmartAssist S2 Set, US; Product Code: 1000100;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2966-2024 Class IIImpella 5.5 SmartAssist Set, JP; Product Code: 1000211;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2967-2024 Class IIImpellaCP SmartAssistSetAPAC, Delo1895; Product Code: 1000302;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2969-2024 Class IIImpella CP Smart Assist Set; Product Code: 1000402;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2970-2024 Class IIImpella 5.5 SmartAssist S2 Set, JP; Product Code: 1000459;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified
Z-2971-2024 Class IIImpella 5.5 with SmartAssist S2 Set, EU; Product Code: 1000482;A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps… 08/05/2024Open, Classified