Recall Z-3006-2026

AVID Medical, Inc. · initiated July 9, 2026

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Halyard TOTAL JOINT Model/Catalog Number: MSBM036-03

Codes / lots: Lot Code: Lot Number: 1632948 UDI: 10809160423578

Reason for recall

Labeling Deficiency: The medical kit labeling fails to identify that the statlock component contains natural rubber latex.

FDA-determined cause: Labeling Change Control

Action taken

On July 9, 2026 Avid Medical ( subsidiary of Owen & Minor) issued a Urgent: Medical Device correction notification to affected consignees. Avid Medical ask consignees to take the following actions: 1. Identify, segregate, and quarantine all impacted product identified. 2. Add warning labels to affected AVID kits to ensure all users are aware of the actions to discard the impacted component prior to use. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. 3. Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. 4. Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. 5. Complete the attached Recall Response Form and return by July 24, 2026.

Details

Recall number
Z-3006-2026
Initiated
July 9, 2026
Posted by FDA
August 21, 2026
Product code
LRO – General Surgery Tray
Quantity
63 kits
Distribution
US Nationwide distribution in the states of MS, TX.
Recalling firm
AVID Medical, Inc., Toano, VA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-3005-2026 Class IIBrand Name: Halyard Product Name: DAVINCI TRUCUSTOM BUNDLE containing TRAY FOLEY 16FR…Labeling Deficiency: The medical kit labeling fails to identify that the statlock component contains natural… 07/09/2026Open, Classified
Z-3007-2026 Class IIHalyard TOTAL JOINT Model/Catalog Number: MSBM036-03Labeling Deficiency: The medical kit labeling fails to identify that the statlock component contains natural… 07/09/2026Open, Classified
Z-3008-2026 Class IIHalyard TOTAL JOINT Model/Catalog Number: MSBM036-03Labeling Deficiency: The medical kit labeling fails to identify that the statlock component contains natural… 07/09/2026Open, Classified
Z-3009-2026 Class IIHalyard TOTAL JOINT Model/Catalog Number: MSBM036-03Labeling Deficiency: The medical kit labeling fails to identify that the statlock component contains natural… 07/09/2026Open, Classified
Z-3010-2026 Class IIHalyard TOTAL JOINT Model/Catalog Number: MSBM036-03Labeling Deficiency: The medical kit labeling fails to identify that the statlock component contains natural… 07/09/2026Open, Classified
Z-3011-2026 Class IIHalyard TOTAL JOINT Model/Catalog Number: MSBM036-03Labeling Deficiency: The medical kit labeling fails to identify that the statlock component contains natural… 07/09/2026Open, Classified