Recall Z-3010-2026
AVID Medical, Inc. · initiated July 9, 2026
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Halyard TOTAL JOINT Model/Catalog Number: MSBM036-03
Codes / lots: Lot Code: Lot Number: 1673233 UDI: 10809160457887
Reason for recall
Labeling Deficiency: The medical kit labeling fails to identify that the statlock component contains natural rubber latex.
FDA-determined cause: Labeling Change Control
Action taken
On July 9, 2026 Avid Medical ( subsidiary of Owen & Minor) issued a Urgent: Medical Device correction notification to affected consignees. Avid Medical ask consignees to take the following actions: 1. Identify, segregate, and quarantine all impacted product identified. 2. Add warning labels to affected AVID kits to ensure all users are aware of the actions to discard the impacted component prior to use. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. 3. Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. 4. Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. 5. Complete the attached Recall Response Form and return by July 24, 2026.
Details
- Recall number
- Z-3010-2026
- Initiated
- July 9, 2026
- Posted by FDA
- August 21, 2026
- Product code
- LRO – General Surgery Tray
- Quantity
- 130 KITS
- Distribution
- US Nationwide distribution in the states of MS, TX.
- Recalling firm
- AVID Medical, Inc., Toano, VA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-3005-2026 Class II | Brand Name: Halyard Product Name: DAVINCI TRUCUSTOM BUNDLE containing TRAY FOLEY 16FR…Labeling Deficiency: The medical kit labeling fails to identify that the statlock component contains natural… | 07/09/2026Open, Classified |
| Z-3006-2026 Class II | Halyard TOTAL JOINT Model/Catalog Number: MSBM036-03Labeling Deficiency: The medical kit labeling fails to identify that the statlock component contains natural… | 07/09/2026Open, Classified |
| Z-3007-2026 Class II | Halyard TOTAL JOINT Model/Catalog Number: MSBM036-03Labeling Deficiency: The medical kit labeling fails to identify that the statlock component contains natural… | 07/09/2026Open, Classified |
| Z-3008-2026 Class II | Halyard TOTAL JOINT Model/Catalog Number: MSBM036-03Labeling Deficiency: The medical kit labeling fails to identify that the statlock component contains natural… | 07/09/2026Open, Classified |
| Z-3009-2026 Class II | Halyard TOTAL JOINT Model/Catalog Number: MSBM036-03Labeling Deficiency: The medical kit labeling fails to identify that the statlock component contains natural… | 07/09/2026Open, Classified |
| Z-3011-2026 Class II | Halyard TOTAL JOINT Model/Catalog Number: MSBM036-03Labeling Deficiency: The medical kit labeling fails to identify that the statlock component contains natural… | 07/09/2026Open, Classified |