Recall Z-0889-2025
GE Medical Systems Israel, Functional Imaging · initiated December 18, 2024
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
GE Healthcare Brivo NM 615, Model/Catalog Numbers: 1) 5376204-15, 2) H2401MA, 3) H3100TA; System, Tomography, Computed, Emission
Codes / lots: 1) 5376204-15: UDI/DI 00840682145503, System ID 502629NMGD, Serial Number NT2UG2100002BH; UDI/DI 00840682117869, System ID 509455BRIVO, Serial Number NT24G1700010BH, UDI/DI N/A, System ID 248544D630, Serial Number NT2QG1900003BH\NT2UG2100001BH. 2) H2401MA: UDI/DI Not Available, System ID 501552D615, Serial Number NT24G1600023; UDI/DI Not Available, System ID 100035454149, Serial Number T2CG15015/100035454149; UDI/DI Not Available, System ID 337643NM, Serial Number NT24G1600021; UDI/DI Not Available, System ID 718963B615, Serial Number NT24G1700009; UDI/DI Not Available, System ID Not Available, Serial Number NT24G1600020; UDI/DI Not Available, System ID 509455BRIVO, Serial Number NT24G1700010; UDI/DI Not Available, System ID 509455BRIVO, Serial Number NT24G1800001; UDI/DI (01)00840682117869(11)17060021NT24G1700003BH, System ID NM6583AR, Serial Number NT24G1700003; UDI/DI (01)00840682117869(11)17090021NT24G1700006BH, System ID NT24G1700006BH, Serial Number NT24G1700006; UDI/DI (01)00840682117869(11)17080021NT24G1700004BH, System ID NT24G1700004BH, Serial Number NT24G1700004; UDI/DI Not Available, System ID Not Available, Serial Number NT24G1700008; UDI/DI Not Available, System ID…
Reason for recall
GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
FDA-determined cause: Package design/selection
Action taken
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 12/18/2024. The notice explained the safety issue and requested the following actions be taken: There are two populations potentially affected by this notification: NOTE: Specific serial numbers of the following affected systems: Infinia Nuclear Medicine Systems: Infinia, Infinia Hawkeye, Infinia II, Infinia II Hawkeye 4; VG Nuclear Medicine systems: Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH; Brivo NM 615; Discovery NM 630; Discovery NM/CT 670, Discovery NM/CT 670 ES, Discovery NM/CT 670 DR, Discovery NM/CT 670 PRO, Discovery NM/CT 670 CZT; Optima NM/CT 640. For Serial Numbers listed in Appendix A of this letter: GE HealthCare requests that you STOP USING the system until GE HealthCare has inspected and if necessary, corrected your system and provided you with a letter indicating the system can be returned to clinical use. GE HealthCare has determined that these systems could have previously been transported or relocated without proper detector support. For Serial Numbers listed in Appendix B of this letter: You can continue to use your system, as GE HealthCare has determined that these systems have not previously been transported or relocated without proper detector support. For Serial Numbers listed in either Appendix A or Appendix B of this letter: GE HealthCare requests that you DO NOT physically relocate or transport your system from its current location. If you need to move or transport your system, please email GE HealthCare at recall.40910@gehealthcare.com to receive specific instructions to follow before moving the system. When emailing this address, please provide the following information: " System serial number " Current system location " The reason for the move or transport, and subsequent destination Do not proceed with a physical relocation or transport of your system until you have received these instructions and taken the actions descri
Details
- Recall number
- Z-0889-2025
- Initiated
- December 18, 2024
- Posted by FDA
- January 10, 2025
- Product code
- KPS – System, Tomography, Computed, Emission
- Quantity
- 116 units
- Distribution
- Worldwide
- Recalling firm
- GE Medical Systems Israel, Functional Imaging, Tirat Carmel
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0888-2025 Class II | GE Healthcare Infinia I, Model/Catalog Numbers: 1) 5178728, 2) H3000WE, 3) H3000WM, 4)…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… | 12/18/2024Open, Classified |
| Z-0890-2025 Class II | GE Healthcare Discovery NM 630, Model/Catalog Numbers: 1) 5376204-30-2, 2) H2401CF, 3)…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… | 12/18/2024Open, Classified |
| Z-0891-2025 Class II | GE Healthcare Discovery NM/CT 670, Model/Catalog Numbers: 1) 5376204-70-54-2, 2)…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… | 12/18/2024Open, Classified |
| Z-0892-2025 Class II | GE Healthcare Discovery NM/CT 670 DR, Model/Catalog Numbers: 1) 5376204-70-56-2; 2)…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… | 12/18/2024Open, Classified |
| Z-0893-2025 Class II | GE Healthcare Discovery NM/CT 670 PRO, Model/Catalog Numbers: 1) 5376204-70-54-2; 2)…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… | 12/18/2024Open, Classified |
| Z-0894-2025 Class II | GE Healthcare Discovery NM/CT 670 ES, Model/Catalog Numbers: 1) H3100XF ; System…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… | 12/18/2024Open, Classified |
| Z-0895-2025 Class II | GE Healthcare Discovery NM/CT 670 CZT, Model/Catalog Numbers: 1) H3906AB; System…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… | 12/18/2024Open, Classified |
| Z-0896-2025 Class II | GE Healthcare Optima NM/CT 640, Model/Catalog Numbers: 1) H3100JZ; System, Tomography…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… | 12/18/2024Open, Classified |
| Z-0897-2025 Class II | GE Healthcare Infinia II, Model/Catalog Numbers: 1) H2799ZP; 2) H3000WT; System…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… | 12/18/2024Open, Classified |
| Z-0898-2025 Class II | GE Healthcare Infinia I Hawkeye 1, Model/Catalog Numbers: 1) H3000WN; 2) H3000YM; 3)…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… | 12/18/2024Open, Classified |
| Z-0899-2025 Class II | GE Healthcare Infinia II Hawkeye 4, Model/Catalog Numbers: 1) H2799PP; 2) H2799PR; 3)…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… | 12/18/2024Open, Classified |
| Z-0900-2025 Class II | GE Healthcare Varicam Millennium VG Discovery VH, Model/Catalog Numbers: 1) ASM000085…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… | 12/18/2024Open, Classified |