Recall Z-0898-2025

GE Medical Systems Israel, Functional Imaging · initiated December 18, 2024

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

GE Healthcare Infinia I Hawkeye 1, Model/Catalog Numbers: 1) H3000WN; 2) H3000YM; 3) H3000YS; System, Tomography, Computed, Emission

Codes / lots: 1) H3000WN: UDI/DI N/A, System ID YN1593, Serial Number 16433. 2) H3000YM: UDI/DI N/A, System ID 602406INF, Serial Number 16012; UDI/DI N/A, System ID 928537INF, Serial Number 16021; UDI/DI N/A, System ID 661633NM1, Serial Number 15013; UDI/DI N/A, System ID 661633NM1, Serial Number 16051; UDI/DI N/A, System ID 619482INFINIA, Serial Number 16007; UDI/DI N/A, System ID 626359INF2, Serial Number 16175; UDI/DI N/A, System ID 626359INF1, Serial Number 16177; UDI/DI N/A, System ID 916983INFN1, Serial Number 16059; UDI/DI N/A, System ID 858554INF1, Serial Number 16169; UDI/DI N/A, System ID 858554INF2, Serial Number 16186; UDI/DI N/A, System ID 626570INF1, Serial Number 16030; UDI/DI N/A, System ID 310543INF1, Serial Number 16006; UDI/DI N/A, System ID 303320IH, Serial Number 16013; UDI/DI N/A, System ID 860350INIF, Serial Number 16065; UDI/DI N/A, System ID 863687INF1, Serial Number 16182; UDI/DI N/A, System ID 941629INF, Serial Number 16026; UDI/DI N/A, System ID 630801INF, Serial Number 16054; UDI/DI N/A, System ID 316962WI, Serial Number 16048; UDI/DI N/A, System ID 270651INF1, Serial Number 16002; UDI/DI N/A, System ID 225765INF1, Serial Number 16070; UDI/DI N/A, System ID…

Reason for recall

GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

FDA-determined cause: Package design/selection

Action taken

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 12/18/2024. The notice explained the safety issue and requested the following actions be taken: There are two populations potentially affected by this notification: NOTE: Specific serial numbers of the following affected systems: Infinia Nuclear Medicine Systems: Infinia, Infinia Hawkeye, Infinia II, Infinia II Hawkeye 4; VG Nuclear Medicine systems: Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH; Brivo NM 615; Discovery NM 630; Discovery NM/CT 670, Discovery NM/CT 670 ES, Discovery NM/CT 670 DR, Discovery NM/CT 670 PRO, Discovery NM/CT 670 CZT; Optima NM/CT 640. For Serial Numbers listed in Appendix A of this letter: GE HealthCare requests that you STOP USING the system until GE HealthCare has inspected and if necessary, corrected your system and provided you with a letter indicating the system can be returned to clinical use. GE HealthCare has determined that these systems could have previously been transported or relocated without proper detector support. For Serial Numbers listed in Appendix B of this letter: You can continue to use your system, as GE HealthCare has determined that these systems have not previously been transported or relocated without proper detector support. For Serial Numbers listed in either Appendix A or Appendix B of this letter: GE HealthCare requests that you DO NOT physically relocate or transport your system from its current location. If you need to move or transport your system, please email GE HealthCare at recall.40910@gehealthcare.com to receive specific instructions to follow before moving the system. When emailing this address, please provide the following information: " System serial number " Current system location " The reason for the move or transport, and subsequent destination Do not proceed with a physical relocation or transport of your system until you have received these instructions and taken the actions descri

Details

Recall number
Z-0898-2025
Initiated
December 18, 2024
Posted by FDA
January 10, 2025
Product code
KPS – System, Tomography, Computed, Emission
Quantity
93 units
Distribution
Worldwide
Recalling firm
GE Medical Systems Israel, Functional Imaging, Tirat Carmel

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0888-2025 Class IIGE Healthcare Infinia I, Model/Catalog Numbers: 1) 5178728, 2) H3000WE, 3) H3000WM, 4)…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… 12/18/2024Open, Classified
Z-0889-2025 Class IIGE Healthcare Brivo NM 615, Model/Catalog Numbers: 1) 5376204-15, 2) H2401MA, 3)…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… 12/18/2024Open, Classified
Z-0890-2025 Class IIGE Healthcare Discovery NM 630, Model/Catalog Numbers: 1) 5376204-30-2, 2) H2401CF, 3)…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… 12/18/2024Open, Classified
Z-0891-2025 Class IIGE Healthcare Discovery NM/CT 670, Model/Catalog Numbers: 1) 5376204-70-54-2, 2)…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… 12/18/2024Open, Classified
Z-0892-2025 Class IIGE Healthcare Discovery NM/CT 670 DR, Model/Catalog Numbers: 1) 5376204-70-56-2; 2)…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… 12/18/2024Open, Classified
Z-0893-2025 Class IIGE Healthcare Discovery NM/CT 670 PRO, Model/Catalog Numbers: 1) 5376204-70-54-2; 2)…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… 12/18/2024Open, Classified
Z-0894-2025 Class IIGE Healthcare Discovery NM/CT 670 ES, Model/Catalog Numbers: 1) H3100XF ; System…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… 12/18/2024Open, Classified
Z-0895-2025 Class IIGE Healthcare Discovery NM/CT 670 CZT, Model/Catalog Numbers: 1) H3906AB; System…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… 12/18/2024Open, Classified
Z-0896-2025 Class IIGE Healthcare Optima NM/CT 640, Model/Catalog Numbers: 1) H3100JZ; System, Tomography…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… 12/18/2024Open, Classified
Z-0897-2025 Class IIGE Healthcare Infinia II, Model/Catalog Numbers: 1) H2799ZP; 2) H3000WT; System…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… 12/18/2024Open, Classified
Z-0899-2025 Class IIGE Healthcare Infinia II Hawkeye 4, Model/Catalog Numbers: 1) H2799PP; 2) H2799PR; 3)…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… 12/18/2024Open, Classified
Z-0900-2025 Class IIGE Healthcare Varicam Millennium VG Discovery VH, Model/Catalog Numbers: 1) ASM000085…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… 12/18/2024Open, Classified