Recall Z-0897-2025
GE Medical Systems Israel, Functional Imaging · initiated December 18, 2024
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
GE Healthcare Infinia II, Model/Catalog Numbers: 1) H2799ZP; 2) H3000WT; System, Tomography, Computed, Emission
Codes / lots: 1) H2799ZP: UDI/DI N/A, System ID A5109665, Serial Number 16611; 2) H3000WT: UDI/DI N/A, System ID 205408NUC, Serial Number 17419; UDI/DI N/A, System ID 256341NRM1, Serial Number 18226; UDI/DI N/A, System ID 256494NM1, Serial Number 17133; UDI/DI N/A, System ID 256543INF1, Serial Number 17182; UDI/DI N/A, System ID 205408INF2, Serial Number 18157; UDI/DI N/A, System ID 256429INF1, Serial Number 17557; UDI/DI N/A, System ID 256327NUC, Serial Number 17900; UDI/DI N/A, System ID 334418INF, Serial Number 17266; UDI/DI N/A, System ID 256386HKNM2, Serial Number 16970; UDI/DI N/A, System ID 205750INF, Serial Number 16802; UDI/DI N/A, System ID 501776INF, Serial Number 17294; UDI/DI N/A, System ID 479463INF2, Serial Number 16900; UDI/DI N/A, System ID 501202INF, Serial Number 16564; UDI/DI N/A, System ID 479967INF, Serial Number 16817; UDI/DI N/A, System ID 479549INF, Serial Number 18319; UDI/DI N/A, System ID 479751INF, Serial Number 16800; UDI/DI N/A, System ID 928214INF, Serial Number 17075; UDI/DI N/A, System ID 520868INF, Serial Number 18289; UDI/DI N/A, System ID 602588INF, Serial Number 17054; UDI/DI N/A, System ID 623406INF1, Serial Number 17064; UDI/DI N/A, System ID 480512INF2…
Reason for recall
GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
FDA-determined cause: Package design/selection
Action taken
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 12/18/2024. The notice explained the safety issue and requested the following actions be taken: There are two populations potentially affected by this notification: NOTE: Specific serial numbers of the following affected systems: Infinia Nuclear Medicine Systems: Infinia, Infinia Hawkeye, Infinia II, Infinia II Hawkeye 4; VG Nuclear Medicine systems: Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH; Brivo NM 615; Discovery NM 630; Discovery NM/CT 670, Discovery NM/CT 670 ES, Discovery NM/CT 670 DR, Discovery NM/CT 670 PRO, Discovery NM/CT 670 CZT; Optima NM/CT 640. For Serial Numbers listed in Appendix A of this letter: GE HealthCare requests that you STOP USING the system until GE HealthCare has inspected and if necessary, corrected your system and provided you with a letter indicating the system can be returned to clinical use. GE HealthCare has determined that these systems could have previously been transported or relocated without proper detector support. For Serial Numbers listed in Appendix B of this letter: You can continue to use your system, as GE HealthCare has determined that these systems have not previously been transported or relocated without proper detector support. For Serial Numbers listed in either Appendix A or Appendix B of this letter: GE HealthCare requests that you DO NOT physically relocate or transport your system from its current location. If you need to move or transport your system, please email GE HealthCare at recall.40910@gehealthcare.com to receive specific instructions to follow before moving the system. When emailing this address, please provide the following information: " System serial number " Current system location " The reason for the move or transport, and subsequent destination Do not proceed with a physical relocation or transport of your system until you have received these instructions and taken the actions descri
Details
- Recall number
- Z-0897-2025
- Initiated
- December 18, 2024
- Posted by FDA
- January 10, 2025
- Product code
- KPS – System, Tomography, Computed, Emission
- Quantity
- 877 units
- Distribution
- Worldwide
- Recalling firm
- GE Medical Systems Israel, Functional Imaging, Tirat Carmel
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0888-2025 Class II | GE Healthcare Infinia I, Model/Catalog Numbers: 1) 5178728, 2) H3000WE, 3) H3000WM, 4)…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… | 12/18/2024Open, Classified |
| Z-0889-2025 Class II | GE Healthcare Brivo NM 615, Model/Catalog Numbers: 1) 5376204-15, 2) H2401MA, 3)…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… | 12/18/2024Open, Classified |
| Z-0890-2025 Class II | GE Healthcare Discovery NM 630, Model/Catalog Numbers: 1) 5376204-30-2, 2) H2401CF, 3)…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… | 12/18/2024Open, Classified |
| Z-0891-2025 Class II | GE Healthcare Discovery NM/CT 670, Model/Catalog Numbers: 1) 5376204-70-54-2, 2)…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… | 12/18/2024Open, Classified |
| Z-0892-2025 Class II | GE Healthcare Discovery NM/CT 670 DR, Model/Catalog Numbers: 1) 5376204-70-56-2; 2)…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… | 12/18/2024Open, Classified |
| Z-0893-2025 Class II | GE Healthcare Discovery NM/CT 670 PRO, Model/Catalog Numbers: 1) 5376204-70-54-2; 2)…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… | 12/18/2024Open, Classified |
| Z-0894-2025 Class II | GE Healthcare Discovery NM/CT 670 ES, Model/Catalog Numbers: 1) H3100XF ; System…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… | 12/18/2024Open, Classified |
| Z-0895-2025 Class II | GE Healthcare Discovery NM/CT 670 CZT, Model/Catalog Numbers: 1) H3906AB; System…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… | 12/18/2024Open, Classified |
| Z-0896-2025 Class II | GE Healthcare Optima NM/CT 640, Model/Catalog Numbers: 1) H3100JZ; System, Tomography…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… | 12/18/2024Open, Classified |
| Z-0898-2025 Class II | GE Healthcare Infinia I Hawkeye 1, Model/Catalog Numbers: 1) H3000WN; 2) H3000YM; 3)…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… | 12/18/2024Open, Classified |
| Z-0899-2025 Class II | GE Healthcare Infinia II Hawkeye 4, Model/Catalog Numbers: 1) H2799PP; 2) H2799PR; 3)…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… | 12/18/2024Open, Classified |
| Z-0900-2025 Class II | GE Healthcare Varicam Millennium VG Discovery VH, Model/Catalog Numbers: 1) ASM000085…GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630… | 12/18/2024Open, Classified |