Recall Z-1004-2014

Zimmer, Inc. · initiated October 11, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Item codes starting with 001930 LAG SCR 12.7D X 55MM SHOR up to 155 MM, LAG SCR 12.7D X 55 MM SHO up to 100MM, LAG SCR 12.7D X 55MM STAN up to 155 MM, LAG SCR 15.8D X 55MM LARG up to 155 MM

Codes / lots: 61484349 62124275 62157179 61077209 61337135 61484350 60555611 60747379 60874833 61142854 61142854R 60325552 60293083 60879683 61284330 61286943 60085527 60436025 60513856 60879684 61284331 61286947 60145167 60186855 60210921 60815940 60879685 60152929 60293084 60483144 60523344 60483145 60523345 60483145R 60440756 60483146 60523346 61142856 60483147 60523347 60536820 60483148 60523350 60483149 60523353 60483150 60523355 60536821 60483151 60483152 60523360 60483153 60523364 60483154 60523365 61067755 60483155 60523369 60132504 60165597 60206470 60293086 60387521 60452004 60488952 60572992 60653075 60671959 60812149 60984699 61004201 61040107 61092997 61099399 61104820 61142860 61224794 61283207 60127012 60140321 60152930 60208488 60265677 60281409 60293087 60323242 60425131 60452005 60488953 60523371 60561146 60653078 60671960 60812150 60848590 61002194 61002195 61026422 61026423 61026424 61083094 61092998 61099402 61142861 61144218 61195105 61195106 61002194R 61002195R 60096566 60118985 60127013 60137590 60165601 60172576 60234054 60254715 60264345 60317521 60385457 60393375 60436028 60483156 60488954 60513859 60538635 60572993 60576792 60585224 60621346 60653083 60671961…

Reason for recall

Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.

FDA-determined cause: Packaging

Action taken

Distributors received a letter identifying the issue and their responsibilities dated November 4, 2013. These responsibilities include quarantining the affected fielded inventory in their possession, providing hospital risk managers with the recall notification letter and retrieving affected fielded inventory and direct sales from the hospitals.

Details

Recall number
Z-1004-2014
Initiated
October 11, 2013
Posted by FDA
February 20, 2014
Terminated
April 14, 2017
Product code
JDO – Device, Fixation, Proximal Femoral, Implant
Quantity
7,044,680 total
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand, and United Kingdom.
Recalling firm
Zimmer, Inc., Warsaw, IN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1001-2014 Class IIMagna-Fx Cannulated Screw Fixation System 7.0 mm. Item codes starting with 0011420…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1002-2014 Class IIHERBERT MINI BONE SCREW 2 and HERBERT/WHIPPLE C ANN BONE. Items starting with 0011500…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1003-2014 Class IIFREE-LOCK LAG SCREW, 12.7, FREE-LOCK LAG SCREW, 15.8, FREE-LOCK SUPRACOND TUBE and…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1005-2014 Class IIVERSA-FX II KEYLESS TUBE, VERSA-FX II KEYLESS SHORT, VERSA-FX II SHORT TUBE PL, VERSA-FX…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1006-2014 Class IIVERSA-FX KEYLESS 90D X 8 and X 11, VERSA-FX KEYLESS 95D X 8, X 11, and X 14, VERSA-FX…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1007-2014 Class IIM/DN R-F IM NAIL 9MM DIA , 10 MM, 11 MM, 12 MM, 13 MM and 14 MM. Item codes starting…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1008-2014 Class IIRECON SCREW LG 5.5MM DIA and M/DN FEM IM NAIL 8MM DIA to 16 MM DIA. Item codes starting…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1009-2014 Class IIINTERLOCKING IM SCREW SML, MED and LRG; M/DN TIBIAL I/M NAIL 6 MM to 15 MM. Some items…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1010-2014 Class IIINTERLOCKING IM SCREW 4.5 and INTERLOCKING IM SCREW LG Some items starting with codes…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1011-2014 Class IIM/DN HUMERAL I/M NAIL. Sizes from 6 MM to 13 MM. Codes starting with 0022541, 002542 and…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1012-2014 Class IIASIA ONE-PIECE LAG SCREWZimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1013-2014 Class IIITST ANTI-ROTATION SCREW and FEM IM NAIL (10MM, 11 MM, 12 MM, 13 MM, 14 MM and 15 MM)…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1014-2014 Class IIONE-PIECE LAG SCREW 11MM. Item codes starting with 00225900.Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1015-2014 Class IIDISTAL MEDIAL FEMORAL CON, DISTAL POSTERIOR/LATERAL, DISTAL SMALL VOLAR RADIUS, PROXIMAL…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1016-2014 Class IIMOORE HIP PROS LONG FENES, MOORE HIP PROS NRW STR FE, MOORE HIP PROS STR FENEST…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1017-2014 Class IIMOORE HIP PROS LONG FENES, MOORE HIP PROS NRW STR FE, MOORE HIP PROS STR FENEST…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1018-2014 Class IIMODULAR AUSTIN MOORE FENE. Item codes starting with 45502001, 455503001 and 455504001Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1019-2014 Class IIBipolar liners and Multipolar cups with the following dimensions in MM (38, 39, 40, 41…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1020-2014 Class IINEXGEN OSTEOTOMY CORT BON. Item codes starting with 0052500Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1021-2014 Class III/B II KNEE DOMED PATELLA, I/B II KNEE FLUTED ROD (1 and 2), MGII KNEE 10MM PATELLA (2,3…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1022-2014 Class IINEXGEN MIS TIBIA PLATE PR NEXGEN TAPER PLUG NEXGEN PRECOAT PEG TIB PL NEXGEN MICRO…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1023-2014 Class IIMODULAR CUP NEUTRAL LINER MODULAR CUP 10 DEGREE LIN MODULAR CUP 20 DEGREE LIN MODULAR…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1024-2014 Class IITRILOGY ACET SHELL 36MM O TRILOGY ACET SHELL 38MM O TRILOGY ACET SHELL 40MM O TRILOGY…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1025-2014 Class IIHGP II ACETABULAR CUP BONZimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1026-2014 Class IIHGP II ACETABULAR CUP MODZimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1027-2014 Class IIPRESSFIT FEM ST FX 16 X 1Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1028-2014 Class IIVERSYS HERITAGE FEM STEMZimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1029-2014 Class II12/14 UNIPOLAR FEMORAL HE VERSYS CEM LD/FX SZ 11X12 VERSYS CEM LD/FX SZ 12X12 VERSYS CEM…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1030-2014 Class IIZIMTRON 6 DEG 22 HEADSZimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1031-2014 Class II12/14 COCR FEMORAL HEAD (2, 3 and 4); 12/14 FEMORAL HEAD (+7X28M and 7X32M)Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated