Recall Z-1023-2014

Zimmer, Inc. · initiated October 11, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

MODULAR CUP NEUTRAL LINER MODULAR CUP 10 DEGREE LIN MODULAR CUP 20 DEGREE LIN MODULAR CUP 7MM OFFSET LI

Codes / lots: 61547919 61954158 61241463 61821673 62188365 61564304 60919205 61666604 60928241 62086340 62296755 60611445 60841757 60863748 61187640 61224727 61261826 62062929 62173370 60550153 60588810 60828784 61018862 61110916 61159634 62296754 60576173 61209715 61564309 60575226 60595581 60623653 60749578 62316407 60611447 61255394 61364450 61382532 60611244 60817929 60863723 60563186 62268497 61547941 62137191 61892712 61901662 60668061 61584777 61059234 61403759 60497031 61303608 61875716 62087621 62138394 61906179 61919744 61853252 60945780 61403762 61868301 62123088 62330868 61552276 60973584 61159666 61209871 61491267 61731038 61906430 62154608 60806194 60940002 61092880 62304925 61598585 61698771 61712665 61902214 62148657 60612251 60658071 60980492 61072064 61160052 61261860 61642713 61722445 62296659 60715034 61018915 61036921 61136634 61179946 61377132 61504792 61564330 61598593 61652828 61662223 61746897 61885964 61902215 62062927 62166311 62166312 62167440 60664534 60611465 61121730 61491274 61542625 62170707 62296657 62304924 61138050 61392960 61509276 61591304 61662225 61751908 61875723 61885967 61902216 62113239 62183216 62308937 62322488 60715040 60715041 62134061 62292055…

Reason for recall

Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.

FDA-determined cause: Packaging

Action taken

Distributors received a letter identifying the issue and their responsibilities dated November 4, 2013. These responsibilities include quarantining the affected fielded inventory in their possession, providing hospital risk managers with the recall notification letter and retrieving affected fielded inventory and direct sales from the hospitals.

Details

Recall number
Z-1023-2014
Initiated
October 11, 2013
Posted by FDA
February 20, 2014
Terminated
April 14, 2017
Product code
LPH – Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Quantity
7,044,680 total
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand, and United Kingdom.
Recalling firm
Zimmer, Inc., Warsaw, IN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1001-2014 Class IIMagna-Fx Cannulated Screw Fixation System 7.0 mm. Item codes starting with 0011420…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1002-2014 Class IIHERBERT MINI BONE SCREW 2 and HERBERT/WHIPPLE C ANN BONE. Items starting with 0011500…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1003-2014 Class IIFREE-LOCK LAG SCREW, 12.7, FREE-LOCK LAG SCREW, 15.8, FREE-LOCK SUPRACOND TUBE and…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1004-2014 Class IIItem codes starting with 001930 LAG SCR 12.7D X 55MM SHOR up to 155 MM, LAG SCR 12.7D X…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1005-2014 Class IIVERSA-FX II KEYLESS TUBE, VERSA-FX II KEYLESS SHORT, VERSA-FX II SHORT TUBE PL, VERSA-FX…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1006-2014 Class IIVERSA-FX KEYLESS 90D X 8 and X 11, VERSA-FX KEYLESS 95D X 8, X 11, and X 14, VERSA-FX…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1007-2014 Class IIM/DN R-F IM NAIL 9MM DIA , 10 MM, 11 MM, 12 MM, 13 MM and 14 MM. Item codes starting…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1008-2014 Class IIRECON SCREW LG 5.5MM DIA and M/DN FEM IM NAIL 8MM DIA to 16 MM DIA. Item codes starting…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1009-2014 Class IIINTERLOCKING IM SCREW SML, MED and LRG; M/DN TIBIAL I/M NAIL 6 MM to 15 MM. Some items…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1010-2014 Class IIINTERLOCKING IM SCREW 4.5 and INTERLOCKING IM SCREW LG Some items starting with codes…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1011-2014 Class IIM/DN HUMERAL I/M NAIL. Sizes from 6 MM to 13 MM. Codes starting with 0022541, 002542 and…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1012-2014 Class IIASIA ONE-PIECE LAG SCREWZimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1013-2014 Class IIITST ANTI-ROTATION SCREW and FEM IM NAIL (10MM, 11 MM, 12 MM, 13 MM, 14 MM and 15 MM)…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1014-2014 Class IIONE-PIECE LAG SCREW 11MM. Item codes starting with 00225900.Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1015-2014 Class IIDISTAL MEDIAL FEMORAL CON, DISTAL POSTERIOR/LATERAL, DISTAL SMALL VOLAR RADIUS, PROXIMAL…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1016-2014 Class IIMOORE HIP PROS LONG FENES, MOORE HIP PROS NRW STR FE, MOORE HIP PROS STR FENEST…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1017-2014 Class IIMOORE HIP PROS LONG FENES, MOORE HIP PROS NRW STR FE, MOORE HIP PROS STR FENEST…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1018-2014 Class IIMODULAR AUSTIN MOORE FENE. Item codes starting with 45502001, 455503001 and 455504001Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1019-2014 Class IIBipolar liners and Multipolar cups with the following dimensions in MM (38, 39, 40, 41…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1020-2014 Class IINEXGEN OSTEOTOMY CORT BON. Item codes starting with 0052500Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1021-2014 Class III/B II KNEE DOMED PATELLA, I/B II KNEE FLUTED ROD (1 and 2), MGII KNEE 10MM PATELLA (2,3…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1022-2014 Class IINEXGEN MIS TIBIA PLATE PR NEXGEN TAPER PLUG NEXGEN PRECOAT PEG TIB PL NEXGEN MICRO…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1024-2014 Class IITRILOGY ACET SHELL 36MM O TRILOGY ACET SHELL 38MM O TRILOGY ACET SHELL 40MM O TRILOGY…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1025-2014 Class IIHGP II ACETABULAR CUP BONZimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1026-2014 Class IIHGP II ACETABULAR CUP MODZimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1027-2014 Class IIPRESSFIT FEM ST FX 16 X 1Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1028-2014 Class IIVERSYS HERITAGE FEM STEMZimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1029-2014 Class II12/14 UNIPOLAR FEMORAL HE VERSYS CEM LD/FX SZ 11X12 VERSYS CEM LD/FX SZ 12X12 VERSYS CEM…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1030-2014 Class IIZIMTRON 6 DEG 22 HEADSZimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1031-2014 Class II12/14 COCR FEMORAL HEAD (2, 3 and 4); 12/14 FEMORAL HEAD (+7X28M and 7X32M)Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated