Recall Z-1016-2014

Zimmer, Inc. · initiated October 11, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

MOORE HIP PROS LONG FENES, MOORE HIP PROS NRW STR FE, MOORE HIP PROS STR FENEST, THOMPSON HIP PROSTHESIS 3, 4, 5 and 6.

Codes / lots: 60484301 60536380 60565715 60753818 60817438 60864746 61003323 61059514 61081364 61121949 61135827 61387138 61783644 365466 60096346 60097010 60259990 60422473 60459752 60515365 60674982 60715057 60768520 60787821 60799626 60799627 61012710 61029791 61029797 61075345 61103552 61103553 61121950 61121951 61135828 61883369 61901575 61938344 369678 369843 60242811 60260004 60317143 60370104 60411929 60411932 60441611 60501819 60501820 60501821 60540465 60553615 60715059 60759283 60759284 60759285 60787991 60799628 60979642 61003325 61012729 61059519 61121952 61121954 61222215 61339786 61398793 61528632 61541580 61565702 61589080 61613237 61921999 61938346 60085561 60096349 60097011 60124640 60134157 60148294 60154764 60154765 60169374 60169375 60187551 60187552 60242814 60260197 60274423 60301862 60313429 60335338 60395245 60395246 60422533 60442559 60484312 60501825 60511856 60553618 60574472 60600313 60609939 60753835 60768538 60768539 60768541 60768542 60787295 60799629 60799630 60990033 61003326 61003327 61012713 61029792 61042227 61059521 61059540 61059541 61121955 61121958 61286077 61293836 61336916 61356578 61365617 61427119 61506293 61550873 61567517 61685104 61708167…

Reason for recall

Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.

FDA-determined cause: Packaging

Action taken

Distributors received a letter identifying the issue and their responsibilities dated November 4, 2013. These responsibilities include quarantining the affected fielded inventory in their possession, providing hospital risk managers with the recall notification letter and retrieving affected fielded inventory and direct sales from the hospitals.

Details

Recall number
Z-1016-2014
Initiated
October 11, 2013
Posted by FDA
February 20, 2014
Terminated
April 14, 2017
Product code
JDG – Prosthesis, Hip, Femoral Component, Cemented, Metal
Quantity
7,044,680 total
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand, and United Kingdom.
Recalling firm
Zimmer, Inc., Warsaw, IN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1001-2014 Class IIMagna-Fx Cannulated Screw Fixation System 7.0 mm. Item codes starting with 0011420…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1002-2014 Class IIHERBERT MINI BONE SCREW 2 and HERBERT/WHIPPLE C ANN BONE. Items starting with 0011500…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1003-2014 Class IIFREE-LOCK LAG SCREW, 12.7, FREE-LOCK LAG SCREW, 15.8, FREE-LOCK SUPRACOND TUBE and…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1004-2014 Class IIItem codes starting with 001930 LAG SCR 12.7D X 55MM SHOR up to 155 MM, LAG SCR 12.7D X…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1005-2014 Class IIVERSA-FX II KEYLESS TUBE, VERSA-FX II KEYLESS SHORT, VERSA-FX II SHORT TUBE PL, VERSA-FX…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1006-2014 Class IIVERSA-FX KEYLESS 90D X 8 and X 11, VERSA-FX KEYLESS 95D X 8, X 11, and X 14, VERSA-FX…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1007-2014 Class IIM/DN R-F IM NAIL 9MM DIA , 10 MM, 11 MM, 12 MM, 13 MM and 14 MM. Item codes starting…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1008-2014 Class IIRECON SCREW LG 5.5MM DIA and M/DN FEM IM NAIL 8MM DIA to 16 MM DIA. Item codes starting…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1009-2014 Class IIINTERLOCKING IM SCREW SML, MED and LRG; M/DN TIBIAL I/M NAIL 6 MM to 15 MM. Some items…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1010-2014 Class IIINTERLOCKING IM SCREW 4.5 and INTERLOCKING IM SCREW LG Some items starting with codes…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1011-2014 Class IIM/DN HUMERAL I/M NAIL. Sizes from 6 MM to 13 MM. Codes starting with 0022541, 002542 and…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1012-2014 Class IIASIA ONE-PIECE LAG SCREWZimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1013-2014 Class IIITST ANTI-ROTATION SCREW and FEM IM NAIL (10MM, 11 MM, 12 MM, 13 MM, 14 MM and 15 MM)…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1014-2014 Class IIONE-PIECE LAG SCREW 11MM. Item codes starting with 00225900.Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1015-2014 Class IIDISTAL MEDIAL FEMORAL CON, DISTAL POSTERIOR/LATERAL, DISTAL SMALL VOLAR RADIUS, PROXIMAL…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1017-2014 Class IIMOORE HIP PROS LONG FENES, MOORE HIP PROS NRW STR FE, MOORE HIP PROS STR FENEST…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1018-2014 Class IIMODULAR AUSTIN MOORE FENE. Item codes starting with 45502001, 455503001 and 455504001Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1019-2014 Class IIBipolar liners and Multipolar cups with the following dimensions in MM (38, 39, 40, 41…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1020-2014 Class IINEXGEN OSTEOTOMY CORT BON. Item codes starting with 0052500Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1021-2014 Class III/B II KNEE DOMED PATELLA, I/B II KNEE FLUTED ROD (1 and 2), MGII KNEE 10MM PATELLA (2,3…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1022-2014 Class IINEXGEN MIS TIBIA PLATE PR NEXGEN TAPER PLUG NEXGEN PRECOAT PEG TIB PL NEXGEN MICRO…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1023-2014 Class IIMODULAR CUP NEUTRAL LINER MODULAR CUP 10 DEGREE LIN MODULAR CUP 20 DEGREE LIN MODULAR…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1024-2014 Class IITRILOGY ACET SHELL 36MM O TRILOGY ACET SHELL 38MM O TRILOGY ACET SHELL 40MM O TRILOGY…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1025-2014 Class IIHGP II ACETABULAR CUP BONZimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1026-2014 Class IIHGP II ACETABULAR CUP MODZimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1027-2014 Class IIPRESSFIT FEM ST FX 16 X 1Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1028-2014 Class IIVERSYS HERITAGE FEM STEMZimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1029-2014 Class II12/14 UNIPOLAR FEMORAL HE VERSYS CEM LD/FX SZ 11X12 VERSYS CEM LD/FX SZ 12X12 VERSYS CEM…Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1030-2014 Class IIZIMTRON 6 DEG 22 HEADSZimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated
Z-1031-2014 Class II12/14 COCR FEMORAL HEAD (2, 3 and 4); 12/14 FEMORAL HEAD (+7X28M and 7X32M)Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V.… 10/11/2013Terminated