Recall Z-2219-2023

Datascope Corp. · initiated June 23, 2023

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

MEGA 7.5Fr.IAB , Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES D684-00-0295-01 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES (APA) D684-00-0295-02 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES (W/O STATLOCK) D684-00-0295-03 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES (CHINA) D684-00-0295-05 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES AND APA (CHINA) D684-00-0295-06 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES W/O STATLOCK - INDIA D684-00-0295-07 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES W/O STATLOCK (APA) - INDIA D684-00-0295-08 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES (TURKEY) D684-00-0295-09 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES (APA) (TURKEY) D684-00-0295-10 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES, US ONLY D684-00-0295-01U MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES APA, US ONLY D684-00-0295-02U MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES D684-00-0294-01 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES (APA) D684-00-0294-02 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES (W/O STATLOCK) D684-00-0294-03 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES (CHINA) D684-00-0294-05 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES AND APA (CHINA) D684-00-0294-06 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES W/O STATLOCK - INDIA D684-00-0294-07 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES W/O STATLOCK (APA) INDIA D684-00-0294-08 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES (TURKEY) D684-00-0294-09 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES (APA) (TURKEY) D684-00-0294-10 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES, US ONLY D684-00-0294-01U MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES APA, US ONLY D684-00-0294-02U PACKAGED INSERTION KIT - MEGA 7.5 Fr. 30/40 cc IABs D884-00-0019-21

Codes / lots: Product Description UDI MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES 10607567107974 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES (APA) 10607567107981 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES (W/O STATLOCK) 10607567109404 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES (CHINA) 10607567109039 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES AND APA (CHINA) 10607567109046 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES W/O STATLOCK - INDIA 10607567111339 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES W/O STATLOCK (APA) - INDIA 10607567111346 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES (TURKEY) 10607567113241 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES (APA) (TURKEY) 10607567113258 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES, US ONLY 10605767109596 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES APA, US ONLY 10605767109602 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES 10607567107950 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES (APA) 10607567107967 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES (W/O STATLOCK) 10607567109398 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES (CHINA) 10607567109015 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES AND APA (CHINA) 10607567109022 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES W/O STATLOCK - INDIA 10607567111315 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES W/O STATLOCK (APA) INDIA…

Reason for recall

During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

FDA-determined cause: Under Investigation by firm

Action taken

Firm notified customers on June 23, 2023 via letter. Customers were made aware of the potential for the device to break during use and the risk of using the device. Customers were given alternate solutions to continue using the IAB Catheter with alternative sheath/dilator combinations. Customers may also return their affected product for a credit. The firm will continue to ship product to customers who choose to continue receiving it. However, product currently being shipped is potentially affected by the same issue. Datascope/Getinge has identified the issue with the introducer dilator and is working with the supplier towards a resolution. If you have any questions, please contact your Datascope/Getinge representative or call Datascope/Getinge Customer Service at 1-888-943-8872, options 2, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).

Details

Recall number
Z-2219-2023
Initiated
June 23, 2023
Posted by FDA
July 21, 2023
Product code
DSP – System, Balloon, Intra-Aortic And Control
Quantity
321,609 total kits
Distribution
Domestic distribution nationwide. International distribution worldwide.
Recalling firm
Datascope Corp., Fairfield, NJ

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2216-2023 Class IIYAMATO PLUS IAB; intra-aortic balloon catheter and accessories used to provide counter…During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the… 06/23/2023Open, Classified
Z-2217-2023 Class IISENSATION 7 Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide…During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the… 06/23/2023Open, Classified
Z-2218-2023 Class IITRANS-RAY IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter…During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the… 06/23/2023Open, Classified
Z-2220-2023 Class IIMEGA 8Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter…During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the… 06/23/2023Open, Classified
Z-2221-2023 Class IISENSATION PLUS 7.5Fr IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide…During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the… 06/23/2023Open, Classified
Z-2222-2023 Class IISENSATION PLUS 8Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide…During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the… 06/23/2023Open, Classified
Z-2223-2023 Class IILINEAR 7.5Fr.IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter…During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the… 06/23/2023Open, Classified
Z-2224-2023 Class IIREINFORCED INTRODUCER SET, Intra-Aortic Balloon Catheter and Accessories, used to…During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the… 06/23/2023Open, Classified