Recall Z-2220-2023
Datascope Corp. · initiated June 23, 2023
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
MEGA 8Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number MEGA 8Fr. 50cc IAB WITH STATLOCK D684-00-0296-01 MEGA 8Fr. 50cc IAB WITH STATLOCK & APA D684-00-0296-02 MEGA 8Fr. 50cc IAB WITH ACCESSORIES (W/O STATLOCK) D684-00-0296-03 MEGA 8Fr. 50cc IAB WITH ACCESSORIES (TURKEY) D684-00-0296-09 MEGA 8Fr. 50cc IAB WITH ACCESSORIES WITH APA (TURKEY) D684-00-0296-10 MEGA 8Fr. 50cc IAB WITH ACCESSORIES, US ONLY D684-00-0296-01U MEGA 8Fr. 50cc IAB WITH ACCESSORIES, (APA) US ONLY D684-00-0296-02U MEGA 8Fr. 50cc IAB WITH ACCESSORIES (CHINA) D684-00-0498-01 MEGA 8Fr. 50cc IAB WITH ACCESSORIES (INDIA) D684-00-0498-07 MEGA 8Fr. 50cc IAB WITH ACCESSORIES (APA) (INDIA) D684-00-0498-08 PACKAGED INSERTION KIT - MEGA 8 Fr. 50 cc IAB D884-00-0019-17
Codes / lots: Product Description UDI MEGA 8Fr. 50cc IAB WITH STATLOCK 10607567108001 MEGA 8Fr. 50cc IAB WITH STATLOCK & APA 10607567108018 MEGA 8Fr. 50cc IAB WITH ACCESSORIES (W/O STATLOCK) 10607567109411 MEGA 8Fr. 50cc IAB WITH ACCESSORIES (TURKEY) 10607567113265 MEGA 8Fr. 50cc IAB WITH ACCESSORIES WITH APA (TURKEY) 10607567113272 MEGA 8Fr. 50cc IAB WITH ACCESSORIES, US ONLY 10607567109619 MEGA 8Fr. 50cc IAB WITH ACCESSORIES, (APA) US ONLY 10607567109626 MEGA 8Fr. 50cc IAB WITH ACCESSORIES (CHINA) 10607567107301 MEGA 8Fr. 50cc IAB WITH ACCESSORIES (INDIA) 10607567109756 MEGA 8Fr. 50cc IAB WITH ACCESSORIES (APA) (INDIA) 10607567109763 PACKAGED INSERTION KIT - MEGA 8 Fr. 50 cc IAB 10607567107493
Reason for recall
During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.
FDA-determined cause: Under Investigation by firm
Action taken
Firm notified customers on June 23, 2023 via letter. Customers were made aware of the potential for the device to break during use and the risk of using the device. Customers were given alternate solutions to continue using the IAB Catheter with alternative sheath/dilator combinations. Customers may also return their affected product for a credit. The firm will continue to ship product to customers who choose to continue receiving it. However, product currently being shipped is potentially affected by the same issue. Datascope/Getinge has identified the issue with the introducer dilator and is working with the supplier towards a resolution. If you have any questions, please contact your Datascope/Getinge representative or call Datascope/Getinge Customer Service at 1-888-943-8872, options 2, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
Details
- Recall number
- Z-2220-2023
- Initiated
- June 23, 2023
- Posted by FDA
- July 21, 2023
- Product code
- DSP – System, Balloon, Intra-Aortic And Control
- Quantity
- 321,609 total kits
- Distribution
- Domestic distribution nationwide. International distribution worldwide.
- Recalling firm
- Datascope Corp., Fairfield, NJ
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2216-2023 Class II | YAMATO PLUS IAB; intra-aortic balloon catheter and accessories used to provide counter…During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the… | 06/23/2023Open, Classified |
| Z-2217-2023 Class II | SENSATION 7 Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide…During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the… | 06/23/2023Open, Classified |
| Z-2218-2023 Class II | TRANS-RAY IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter…During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the… | 06/23/2023Open, Classified |
| Z-2219-2023 Class II | MEGA 7.5Fr.IAB , Intra-Aortic Balloon Catheter and Accessories, used to provide counter…During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the… | 06/23/2023Open, Classified |
| Z-2221-2023 Class II | SENSATION PLUS 7.5Fr IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide…During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the… | 06/23/2023Open, Classified |
| Z-2222-2023 Class II | SENSATION PLUS 8Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide…During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the… | 06/23/2023Open, Classified |
| Z-2223-2023 Class II | LINEAR 7.5Fr.IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter…During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the… | 06/23/2023Open, Classified |
| Z-2224-2023 Class II | REINFORCED INTRODUCER SET, Intra-Aortic Balloon Catheter and Accessories, used to…During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the… | 06/23/2023Open, Classified |