Recall Z-2223-2023
Datascope Corp. · initiated June 23, 2023
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
LINEAR 7.5Fr.IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES D684-00-0479-01 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (APA) D684-00-0479-02 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (CHINA) D684-00-0479-01C LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (APA)(CHINA) D684-00-0479-02C LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (INDIA) D684-00-0479-07 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (APA) (INDIA) D684-00-0479-08 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (TURKEY) D684-00-0479-09 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES WITH APA (TURKEY) D684-00-0479-10 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES, US ONLY D684-00-0479-01U LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES, (APA) US ONLY D684-00-0479-02U LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES D684-00-0480-01 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (APA) D684-00-0480-02 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (CHINA) D684-00-0480-01C LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (APA)(CHINA) D684-00-0480-02C LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (INDIA) D684-00-0480-07 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (APA)(INDIA) D684-00-0480-08 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (TURKEY) D684-00-0480-09 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES WITH APA (TURKEY) D684-00-0480-10 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES, US ONLY D684-00-0480-01U LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES, (APA) US ONLY D684-00-0480-02U LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES D684-00-0478-01 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (APA) D684-00-0478-02 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (CHINA) D684-00-0478-01C LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (APA) (CHINA) D684-00-0478-02C LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (INDIA) D684-00-0478-07 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (APA) (INDIA) D684-00-0478-08 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (TURKEY) D684-00-0478-09 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES WITH APA (TURKEY) D684-00-0478-10 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES, US ONLY D684-00-0478-01U LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (APA) US ONLY D684-00-0478-02U PACKAGED INSERTION KIT - LINEAR 7.5 Fr. 25 cc IABs D884-00-0019-12 PACKAGED INSERTION KIT - LINEAR 7.5 Fr. 34/40 cc IABs D884-00-0019-13
Codes / lots: Product Description UDI LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES 10605767106557 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (APA) 10605767106540 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (CHINA) 10605767109329 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (APA)(CHINA) 10605767109336 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (INDIA) 10605767109695 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (APA) (INDIA) 10605767109701 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (TURKEY) 10605767113180 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES WITH APA (TURKEY) 10605767113142 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES, US ONLY 10605767109503 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES, (APA) US ONLY 10605767109510 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES 10607567106564 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (APA) 10607567106571 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (CHINA) 10607567109343 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (APA)(CHINA) 10607567109350 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (INDIA) 10607567109718 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (APA)(INDIA) 10607567109725 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (TURKEY) 10607567113197 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES WITH APA (TURKEY) 10607567113159 LINEAR…
Reason for recall
During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.
FDA-determined cause: Under Investigation by firm
Action taken
Firm notified customers on June 23, 2023 via letter. Customers were made aware of the potential for the device to break during use and the risk of using the device. Customers were given alternate solutions to continue using the IAB Catheter with alternative sheath/dilator combinations. Customers may also return their affected product for a credit. The firm will continue to ship product to customers who choose to continue receiving it. However, product currently being shipped is potentially affected by the same issue. Datascope/Getinge has identified the issue with the introducer dilator and is working with the supplier towards a resolution. If you have any questions, please contact your Datascope/Getinge representative or call Datascope/Getinge Customer Service at 1-888-943-8872, options 2, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
Details
- Recall number
- Z-2223-2023
- Initiated
- June 23, 2023
- Posted by FDA
- July 21, 2023
- Product code
- DSP – System, Balloon, Intra-Aortic And Control
- Quantity
- 321,609 total kits
- Distribution
- Domestic distribution nationwide. International distribution worldwide.
- Recalling firm
- Datascope Corp., Fairfield, NJ
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2216-2023 Class II | YAMATO PLUS IAB; intra-aortic balloon catheter and accessories used to provide counter…During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the… | 06/23/2023Open, Classified |
| Z-2217-2023 Class II | SENSATION 7 Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide…During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the… | 06/23/2023Open, Classified |
| Z-2218-2023 Class II | TRANS-RAY IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter…During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the… | 06/23/2023Open, Classified |
| Z-2219-2023 Class II | MEGA 7.5Fr.IAB , Intra-Aortic Balloon Catheter and Accessories, used to provide counter…During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the… | 06/23/2023Open, Classified |
| Z-2220-2023 Class II | MEGA 8Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter…During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the… | 06/23/2023Open, Classified |
| Z-2221-2023 Class II | SENSATION PLUS 7.5Fr IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide…During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the… | 06/23/2023Open, Classified |
| Z-2222-2023 Class II | SENSATION PLUS 8Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide…During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the… | 06/23/2023Open, Classified |
| Z-2224-2023 Class II | REINFORCED INTRODUCER SET, Intra-Aortic Balloon Catheter and Accessories, used to…During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the… | 06/23/2023Open, Classified |