Recall Z-0020-2019
Oridion Medical 1987 Ltd. · initiated July 31, 2018
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
CAPNOSTREAM 20 (US) N, Part Number CS08651
Codes / lots: B200001468 B200001473 B200001476 B200001594 B200001951 B200001980 B200001992 B200002054 B200002129 B200002144 B200002147 B200002149 B200002164 B200002171 B200002204 B200002154 B200001956 B200002085 B200002110 B200001948 B200002048 B200002167 B200002168 B200001953 B200001955 B200001961 B200001964 B200001966 B200001968 B200001969 B200001972 B200001981 B200001991 B200002001 B200002006 B200002008 B200002022 B200002024 B200002033 B200002034 B200002074 B200002075 B200002080 B200002095 B200002111 B200002126 B200001985 B200001717 B200001422 B200001505 B200001599 B200001691 B200001696 B200001783 B200001889 B200001916 B200001919 B200001923 B200002115 B400002214 B400002220 B400002222 B400002223 B200001373 B200001430 B200001480 B200001492 B200001499 B200001534 B200001707 B200001711 B200001720 B200001767 B200001769 B200001868 B200001882 B200001885 B200001887 B200001924 B200001928 B200001935 B200001936 B200001937 B200001940 B200001954 B200001957 B200001973 B200001974 B200001978 B200001983 B200001987 B200001990 B200001993 B200001994 B200001934 B200001939 B200001960 B200001967 B200002003 B200002005 B200002047 B200001496 B200001848 B200002056 B200002061 B200002156 B200001439 B200001449 B200001470…
Reason for recall
The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory default settings when the monitor is powered off.
FDA-determined cause: Nonconforming Material/Component
Action taken
Urgent Medical Device Correction notification letters dated 7/31/18 were distributed to customers. The letters included the following instructions for customers: Share this notification with all care environments where the Capnostream" 20 and Capnostream" 20p bedside patient monitors are used, particularly in the areas where nurse call may be enabled (general care floors). If your facility has distributed these bedside patient monitors to other persons or facilities, please promptly forward a copy of this letter to those recipients. Medtronic is recommending continued use of Capnostream" devices. Users should confirm that the date/time is accurately displayed. If the date/time requires reset, all user defined settings, such as alarms and nurse call, will also require reset until the software is updated. Please follow the instruction in the Operators Manual to set user-defined default settings. The Operators Manual is available at http://www.medtronic.com/covidien/enus/support/product-manuals.html, refer to the Institutional Settings section. Return the completed Acknowledgement and Receipt Form by fax or email even if you have no inventory. In October 2018, download the software update from the Medtronic website via the following link which includes directions on how to download and install the software: http://www.medtronic.com/covidien/en-us/support/patient-monitoring-equipmentsoftware-upgrades.html.
Details
- Recall number
- Z-0020-2019
- Initiated
- July 31, 2018
- Terminated
- September 20, 2022
- Product code
- DQA – Oximeter
- Quantity
- 3143
- Distribution
- The products were distributed US nationwide and Puerto Rico. The products were distributed to the following foreign countries: Argentina, Australia, Austria, Bahrain, Belgium, Canada, Columbia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Kazakhstan, Kuwait, Mexico, Netherlands, Norway, Oman, Pakistan, Panama, Peru…
- Recalling firm
- Oridion Medical 1987 Ltd., Jerusalem, Har Hotzvi
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0021-2019 Class II | CAPNOSTREAM 20P (US) N, Part Number CS08651-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0022-2019 Class II | CAPNOSTREAM 20P (US) N - ROHS, Part Number CS08651-02The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0023-2019 Class II | CAPNOSTREAM 20P (US) N - ROHS, Part Number CS08651-03The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0024-2019 Class II | CAPNOSTREAM 20 (US) N W/PRINTER, Part Number CS08657The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0025-2019 Class II | CAPNOSTREAM 20P (US) N W/PRINTER - ROHS, Part Number CS08657-02The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0026-2019 Class II | CAPNOSTREAM 20P (US) N W/PRINTER, Part Number CS08657-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0027-2019 Class II | CAPNOSTREAM 20P (US) N W/PRINTER - ROHS, Part Number CS08657-03The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0028-2019 Class II | Capnostream20 (US) M, Part Number CS08653The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0029-2019 Class II | Capnostream20 (US) M W/PRINTER, Part Number CS08659The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0030-2019 Class II | Capnostream20 (US) M TO MASIMO, Part Number CS08750The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0031-2019 Class II | Capnostream20 (US) M W/PRINTER TO MASIMO, Part Number CS08751The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0032-2019 Class II | Capnostream20 (US) N REFURBISHED, Part Number CS78651The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0033-2019 Class II | Capnostream20 (US)M REFURBISHED, Part Number CS78653The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0034-2019 Class II | Capnostream20 (US) N W/PRINTER REFURBISHED, Part Number CS78657The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0035-2019 Class II | Capnostream20 (US) M W/PRINTER REFURBISHED, Part Number CS78659The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0036-2019 Class II | CAPNOSTREAM 20 (INTL) N, Part Number CS08652The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0037-2019 Class II | CAPNOSTREAM 20P (INTL) N, Part Number CS08652-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0038-2019 Class II | CAPNOSTREAM 20P (INTL) N - ROHS, Part Number CS08652-02The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0039-2019 Class II | CAPNOSTREAM 20P (INTL) N - ROHS, Part Number CS08652-03The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0040-2019 Class II | CAPNOSTREAM 20 (INTL) N W/PRINTER, Part Number CS08658The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0041-2019 Class II | CAPNOSTREAM 20P (INTL) N W/PRINTER, Part Number CS08658-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0042-2019 Class II | CAPNOSTREAM 20P (INTL) N W/PRINTER - ROHS, CS08658-02The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0043-2019 Class II | CAPNOSTREAM 20P (INTL) N W/PRINTER - ROHS, CS08658-03The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0044-2019 Class II | Capnostream20 (INTL) M, CS08654The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0045-2019 Class II | Capnostream20 M ADULT ICU (CPM), CS08654RAThe date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0046-2019 Class II | Capnostream20 M NICU (CPM), CS08654RNThe date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0047-2019 Class II | Capnostream20 (INTL) M W/PRINTER, CS08660The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0048-2019 Class II | Capnostream20 M W/PRINTER CYVITAL, CS08796The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0049-2019 Class II | Capnostream20 N W/PRINTER CYVITAL, CS08798The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0050-2019 Class II | Capnostream20 N CYVITAL, CS08799The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |