Recall Z-0039-2019
Oridion Medical 1987 Ltd. · initiated July 31, 2018
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
CAPNOSTREAM 20P (INTL) N - ROHS, Part Number CS08652-03
Codes / lots: PA12170508 PA12170656 PA12170657 PA12170702 PA13170746 PA13170769 PA13170777 PA13170782 PA13170783 PA13170786 PA13170787 PA13170798 PA14170801 PA14170802 PA14170803 PA14170807 PA14170808 PA14170811 PA14170812 PA14170813 PA14170814 PA14170815 PA14170816 PA14170817 PA14170818 PA14170819 PA14170820 PA14170821 PA14170822 PA14170823 PA14170824 PA14170825 PA14170826 PA14170827 PA14170828 PA14170829 PA14170830 PA14170831 PA14170832 PA14170833 PA14170834 PA14170835 PA14170836 PA14170837 PA14170838 PA14170839 PA14170840 PA14170841 PA14170842 PA14170843 PA14170844 PA14170845 PA14170846 PA14170847 PA14170848 PA14170849 PA14170850 PA14170851 PA14170852 PA14170853 PA14170854 PA14170855 PA14170856 PA14170857 PA14170858 PA14170859 PA14170860 PA14170861 PA14170862 PA14170863 PA14170864 PA14170865 PA14170866 PA14170867 PA14170868 PA14170869 PA14170870 PA14170871 PA14170872 PA14170873 PA14170874 PA14170875 PA14170876 PA14170877 PA14170878 PA14170879 PA15171049 PA15171050 PA15171051 PA15171053 PA15171054 PA15171055 PA15171056 PA15171057 PA15171058 PA15171059 PA22172143 PA23172302 PA23172307 PA23172308 PA23172310 PA23172311 PA23172312 PA23172313 PA23172314 PA23172315 PA23172316 PA23172317 PA23172318…
Reason for recall
The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory default settings when the monitor is powered off.
FDA-determined cause: Nonconforming Material/Component
Action taken
Urgent Medical Device Correction notification letters dated 7/31/18 were distributed to customers. The letters included the following instructions for customers: Share this notification with all care environments where the Capnostream" 20 and Capnostream" 20p bedside patient monitors are used, particularly in the areas where nurse call may be enabled (general care floors). If your facility has distributed these bedside patient monitors to other persons or facilities, please promptly forward a copy of this letter to those recipients. Medtronic is recommending continued use of Capnostream" devices. Users should confirm that the date/time is accurately displayed. If the date/time requires reset, all user defined settings, such as alarms and nurse call, will also require reset until the software is updated. Please follow the instruction in the Operators Manual to set user-defined default settings. The Operators Manual is available at http://www.medtronic.com/covidien/enus/support/product-manuals.html, refer to the Institutional Settings section. Return the completed Acknowledgement and Receipt Form by fax or email even if you have no inventory. In October 2018, download the software update from the Medtronic website via the following link which includes directions on how to download and install the software: http://www.medtronic.com/covidien/en-us/support/patient-monitoring-equipmentsoftware-upgrades.html.
Details
- Recall number
- Z-0039-2019
- Initiated
- July 31, 2018
- Terminated
- September 20, 2022
- Product code
- DQA – Oximeter
- Quantity
- 264
- Distribution
- The products were distributed US nationwide and Puerto Rico. The products were distributed to the following foreign countries: Argentina, Australia, Austria, Bahrain, Belgium, Canada, Columbia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Kazakhstan, Kuwait, Mexico, Netherlands, Norway, Oman, Pakistan, Panama, Peru…
- Recalling firm
- Oridion Medical 1987 Ltd., Jerusalem, Har Hotzvi
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0020-2019 Class II | CAPNOSTREAM 20 (US) N, Part Number CS08651The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0021-2019 Class II | CAPNOSTREAM 20P (US) N, Part Number CS08651-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0022-2019 Class II | CAPNOSTREAM 20P (US) N - ROHS, Part Number CS08651-02The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0023-2019 Class II | CAPNOSTREAM 20P (US) N - ROHS, Part Number CS08651-03The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0024-2019 Class II | CAPNOSTREAM 20 (US) N W/PRINTER, Part Number CS08657The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0025-2019 Class II | CAPNOSTREAM 20P (US) N W/PRINTER - ROHS, Part Number CS08657-02The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0026-2019 Class II | CAPNOSTREAM 20P (US) N W/PRINTER, Part Number CS08657-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0027-2019 Class II | CAPNOSTREAM 20P (US) N W/PRINTER - ROHS, Part Number CS08657-03The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0028-2019 Class II | Capnostream20 (US) M, Part Number CS08653The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0029-2019 Class II | Capnostream20 (US) M W/PRINTER, Part Number CS08659The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0030-2019 Class II | Capnostream20 (US) M TO MASIMO, Part Number CS08750The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0031-2019 Class II | Capnostream20 (US) M W/PRINTER TO MASIMO, Part Number CS08751The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0032-2019 Class II | Capnostream20 (US) N REFURBISHED, Part Number CS78651The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0033-2019 Class II | Capnostream20 (US)M REFURBISHED, Part Number CS78653The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0034-2019 Class II | Capnostream20 (US) N W/PRINTER REFURBISHED, Part Number CS78657The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0035-2019 Class II | Capnostream20 (US) M W/PRINTER REFURBISHED, Part Number CS78659The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0036-2019 Class II | CAPNOSTREAM 20 (INTL) N, Part Number CS08652The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0037-2019 Class II | CAPNOSTREAM 20P (INTL) N, Part Number CS08652-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0038-2019 Class II | CAPNOSTREAM 20P (INTL) N - ROHS, Part Number CS08652-02The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0040-2019 Class II | CAPNOSTREAM 20 (INTL) N W/PRINTER, Part Number CS08658The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0041-2019 Class II | CAPNOSTREAM 20P (INTL) N W/PRINTER, Part Number CS08658-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0042-2019 Class II | CAPNOSTREAM 20P (INTL) N W/PRINTER - ROHS, CS08658-02The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0043-2019 Class II | CAPNOSTREAM 20P (INTL) N W/PRINTER - ROHS, CS08658-03The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0044-2019 Class II | Capnostream20 (INTL) M, CS08654The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0045-2019 Class II | Capnostream20 M ADULT ICU (CPM), CS08654RAThe date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0046-2019 Class II | Capnostream20 M NICU (CPM), CS08654RNThe date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0047-2019 Class II | Capnostream20 (INTL) M W/PRINTER, CS08660The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0048-2019 Class II | Capnostream20 M W/PRINTER CYVITAL, CS08796The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0049-2019 Class II | Capnostream20 N W/PRINTER CYVITAL, CS08798The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0050-2019 Class II | Capnostream20 N CYVITAL, CS08799The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |