Recall Z-0031-2019

Oridion Medical 1987 Ltd. · initiated July 31, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Capnostream20 (US) M W/PRINTER TO MASIMO, Part Number CS08751

Codes / lots: B500002176 B500002177 B500002186 B500002191 B500002192 B500002194 B500002199 B500002201 B500002202 B500002204 B500002205 B500002206 B500002207 B500002208 B500002209 B500002210 B500002211 B500002212 B500002213 B500002214 B500002215 B500002216 B500002217 B500002218 B500002219 B500002220 B500002224 B500002225 B500002228 B500002229 B500002233 B500002236 B500002239 B500002240 B500002241 B500002245 B500002249 B500002250 B500002251 B500002252 B500002256 B500002258 B500002259 B500002261 B500002262 B500002265 B500002269 B500002270 B500002271 B500002272 B500002273 B500002274 B500002276 B500002277 B500002278 B500002279 B500002280 B500002281 B500002282 B500002283 B500002284 B500002285 B500002286 B500002287 B500002288 B500002291 B500002292 B500002293 B500002294 B500002295 B500002297 B500002298 B500002300 B500002301 B500002170 B500002234 B500002260 B500002266 B500002267 B500002296 B500002350 B500002351 B500002352 B500002353 B500002355 B500002356 B500002358 B500002359 B500002360 B500002361 B500002362 B500002363 B500002365 B500002366 B500002369 B500002371 B500002372 B500002395 B500002409 B500002410 B500002413 B500002414 B500002417 B500002419 B500002420 B500002421 B500002422 B500002423 B500002424…

Reason for recall

The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory default settings when the monitor is powered off.

FDA-determined cause: Nonconforming Material/Component

Action taken

Urgent Medical Device Correction notification letters dated 7/31/18 were distributed to customers. The letters included the following instructions for customers: Share this notification with all care environments where the Capnostream" 20 and Capnostream" 20p bedside patient monitors are used, particularly in the areas where nurse call may be enabled (general care floors). If your facility has distributed these bedside patient monitors to other persons or facilities, please promptly forward a copy of this letter to those recipients. Medtronic is recommending continued use of Capnostream" devices. Users should confirm that the date/time is accurately displayed. If the date/time requires reset, all user defined settings, such as alarms and nurse call, will also require reset until the software is updated. Please follow the instruction in the Operators Manual to set user-defined default settings. The Operators Manual is available at http://www.medtronic.com/covidien/enus/support/product-manuals.html, refer to the Institutional Settings section. Return the completed Acknowledgement and Receipt Form by fax or email even if you have no inventory. In October 2018, download the software update from the Medtronic website via the following link which includes directions on how to download and install the software: http://www.medtronic.com/covidien/en-us/support/patient-monitoring-equipmentsoftware-upgrades.html.

Details

Recall number
Z-0031-2019
Initiated
July 31, 2018
Terminated
September 20, 2022
Product code
DQA – Oximeter
Quantity
400
Distribution
The products were distributed US nationwide and Puerto Rico. The products were distributed to the following foreign countries: Argentina, Australia, Austria, Bahrain, Belgium, Canada, Columbia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Kazakhstan, Kuwait, Mexico, Netherlands, Norway, Oman, Pakistan, Panama, Peru…
Recalling firm
Oridion Medical 1987 Ltd., Jerusalem, Har Hotzvi

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0020-2019 Class IICAPNOSTREAM 20 (US) N, Part Number CS08651The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0021-2019 Class IICAPNOSTREAM 20P (US) N, Part Number CS08651-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0022-2019 Class IICAPNOSTREAM 20P (US) N - ROHS, Part Number CS08651-02The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0023-2019 Class IICAPNOSTREAM 20P (US) N - ROHS, Part Number CS08651-03The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0024-2019 Class IICAPNOSTREAM 20 (US) N W/PRINTER, Part Number CS08657The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0025-2019 Class IICAPNOSTREAM 20P (US) N W/PRINTER - ROHS, Part Number CS08657-02The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0026-2019 Class IICAPNOSTREAM 20P (US) N W/PRINTER, Part Number CS08657-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0027-2019 Class IICAPNOSTREAM 20P (US) N W/PRINTER - ROHS, Part Number CS08657-03The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0028-2019 Class IICapnostream20 (US) M, Part Number CS08653The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0029-2019 Class IICapnostream20 (US) M W/PRINTER, Part Number CS08659The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0030-2019 Class IICapnostream20 (US) M TO MASIMO, Part Number CS08750The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0032-2019 Class IICapnostream20 (US) N REFURBISHED, Part Number CS78651The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0033-2019 Class IICapnostream20 (US)M REFURBISHED, Part Number CS78653The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0034-2019 Class IICapnostream20 (US) N W/PRINTER REFURBISHED, Part Number CS78657The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0035-2019 Class IICapnostream20 (US) M W/PRINTER REFURBISHED, Part Number CS78659The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0036-2019 Class IICAPNOSTREAM 20 (INTL) N, Part Number CS08652The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0037-2019 Class IICAPNOSTREAM 20P (INTL) N, Part Number CS08652-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0038-2019 Class IICAPNOSTREAM 20P (INTL) N - ROHS, Part Number CS08652-02The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0039-2019 Class IICAPNOSTREAM 20P (INTL) N - ROHS, Part Number CS08652-03The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0040-2019 Class IICAPNOSTREAM 20 (INTL) N W/PRINTER, Part Number CS08658The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0041-2019 Class IICAPNOSTREAM 20P (INTL) N W/PRINTER, Part Number CS08658-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0042-2019 Class IICAPNOSTREAM 20P (INTL) N W/PRINTER - ROHS, CS08658-02The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0043-2019 Class IICAPNOSTREAM 20P (INTL) N W/PRINTER - ROHS, CS08658-03The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0044-2019 Class IICapnostream20 (INTL) M, CS08654The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0045-2019 Class IICapnostream20 M ADULT ICU (CPM), CS08654RAThe date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0046-2019 Class IICapnostream20 M NICU (CPM), CS08654RNThe date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0047-2019 Class IICapnostream20 (INTL) M W/PRINTER, CS08660The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0048-2019 Class IICapnostream20 M W/PRINTER CYVITAL, CS08796The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0049-2019 Class IICapnostream20 N W/PRINTER CYVITAL, CS08798The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0050-2019 Class IICapnostream20 N CYVITAL, CS08799The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated