Recall Z-0042-2019
Oridion Medical 1987 Ltd. · initiated July 31, 2018
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
CAPNOSTREAM 20P (INTL) N W/PRINTER - ROHS, CS08658-02
Codes / lots: B819140013 B819140014 B819140015 B819140024 B819140030 B819140042 B819140044 B819140049 B820140072 B820140075 B820140077 B820140078 B820140079 B820140080 B820140081 B819140003 B819140005 B819140010 B819140016 B819140018 B819140023 B819140025 B819140028 B819140029 B819140035 B819140036 B819140039 B819140047 B819140056 B819140059 B819140062 B820140067 B820140068 B820140071 B820140073 B820140074 B820140082 B819140009 B819140022 B819140027 B819140040 B819140041 B819140046 B819140053 B819140055 B819140057 B819140058 B819140061 B819140064 B819140066 B820140070 B839141804 B839141806 B839141810 B840141817 B840141841 B840141842 B840141836 B840141840 B840141845 B840141846 B848143029 B850143387 B851143493 B851143500 B851143504 B851143505 B851143534 B851143541 B851143548 B851143550 B851143553 B851143556 B851143566 B851143569 B851143578 B802153887 B802153892 B802153936 B802153947 B803153953 B802153812 B802153817 B802153834 B802153859 B802153879 B802153891 B802153903 B802153914 B802153940 B802153942 B803154038 B803154050 B804154170 B804154182 B804154191 B804154193 B804154199 B804154216 B804154245 B805154263 B805154265 B805154267 B851143580 B851143609 B851143638 B811155297 B811155312 B811155316…
Reason for recall
The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory default settings when the monitor is powered off.
FDA-determined cause: Nonconforming Material/Component
Action taken
Urgent Medical Device Correction notification letters dated 7/31/18 were distributed to customers. The letters included the following instructions for customers: Share this notification with all care environments where the Capnostream" 20 and Capnostream" 20p bedside patient monitors are used, particularly in the areas where nurse call may be enabled (general care floors). If your facility has distributed these bedside patient monitors to other persons or facilities, please promptly forward a copy of this letter to those recipients. Medtronic is recommending continued use of Capnostream" devices. Users should confirm that the date/time is accurately displayed. If the date/time requires reset, all user defined settings, such as alarms and nurse call, will also require reset until the software is updated. Please follow the instruction in the Operators Manual to set user-defined default settings. The Operators Manual is available at http://www.medtronic.com/covidien/enus/support/product-manuals.html, refer to the Institutional Settings section. Return the completed Acknowledgement and Receipt Form by fax or email even if you have no inventory. In October 2018, download the software update from the Medtronic website via the following link which includes directions on how to download and install the software: http://www.medtronic.com/covidien/en-us/support/patient-monitoring-equipmentsoftware-upgrades.html.
Details
- Recall number
- Z-0042-2019
- Initiated
- July 31, 2018
- Terminated
- September 20, 2022
- Product code
- DQA – Oximeter
- Quantity
- 383
- Distribution
- The products were distributed US nationwide and Puerto Rico. The products were distributed to the following foreign countries: Argentina, Australia, Austria, Bahrain, Belgium, Canada, Columbia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Kazakhstan, Kuwait, Mexico, Netherlands, Norway, Oman, Pakistan, Panama, Peru…
- Recalling firm
- Oridion Medical 1987 Ltd., Jerusalem, Har Hotzvi
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-0020-2019 Class II | CAPNOSTREAM 20 (US) N, Part Number CS08651The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0021-2019 Class II | CAPNOSTREAM 20P (US) N, Part Number CS08651-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0022-2019 Class II | CAPNOSTREAM 20P (US) N - ROHS, Part Number CS08651-02The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0023-2019 Class II | CAPNOSTREAM 20P (US) N - ROHS, Part Number CS08651-03The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0024-2019 Class II | CAPNOSTREAM 20 (US) N W/PRINTER, Part Number CS08657The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0025-2019 Class II | CAPNOSTREAM 20P (US) N W/PRINTER - ROHS, Part Number CS08657-02The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0026-2019 Class II | CAPNOSTREAM 20P (US) N W/PRINTER, Part Number CS08657-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0027-2019 Class II | CAPNOSTREAM 20P (US) N W/PRINTER - ROHS, Part Number CS08657-03The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0028-2019 Class II | Capnostream20 (US) M, Part Number CS08653The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0029-2019 Class II | Capnostream20 (US) M W/PRINTER, Part Number CS08659The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0030-2019 Class II | Capnostream20 (US) M TO MASIMO, Part Number CS08750The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0031-2019 Class II | Capnostream20 (US) M W/PRINTER TO MASIMO, Part Number CS08751The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0032-2019 Class II | Capnostream20 (US) N REFURBISHED, Part Number CS78651The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0033-2019 Class II | Capnostream20 (US)M REFURBISHED, Part Number CS78653The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0034-2019 Class II | Capnostream20 (US) N W/PRINTER REFURBISHED, Part Number CS78657The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0035-2019 Class II | Capnostream20 (US) M W/PRINTER REFURBISHED, Part Number CS78659The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0036-2019 Class II | CAPNOSTREAM 20 (INTL) N, Part Number CS08652The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0037-2019 Class II | CAPNOSTREAM 20P (INTL) N, Part Number CS08652-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0038-2019 Class II | CAPNOSTREAM 20P (INTL) N - ROHS, Part Number CS08652-02The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0039-2019 Class II | CAPNOSTREAM 20P (INTL) N - ROHS, Part Number CS08652-03The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0040-2019 Class II | CAPNOSTREAM 20 (INTL) N W/PRINTER, Part Number CS08658The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0041-2019 Class II | CAPNOSTREAM 20P (INTL) N W/PRINTER, Part Number CS08658-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0043-2019 Class II | CAPNOSTREAM 20P (INTL) N W/PRINTER - ROHS, CS08658-03The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0044-2019 Class II | Capnostream20 (INTL) M, CS08654The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0045-2019 Class II | Capnostream20 M ADULT ICU (CPM), CS08654RAThe date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0046-2019 Class II | Capnostream20 M NICU (CPM), CS08654RNThe date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0047-2019 Class II | Capnostream20 (INTL) M W/PRINTER, CS08660The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0048-2019 Class II | Capnostream20 M W/PRINTER CYVITAL, CS08796The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0049-2019 Class II | Capnostream20 N W/PRINTER CYVITAL, CS08798The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |
| Z-0050-2019 Class II | Capnostream20 N CYVITAL, CS08799The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… | 07/31/2018Terminated |