Recall Z-0043-2019

Oridion Medical 1987 Ltd. · initiated July 31, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

CAPNOSTREAM 20P (INTL) N W/PRINTER - ROHS, CS08658-03

Codes / lots: PA14170895 PA14170897 PA14170903 PA14170930 PA14170937 PA14170944 PA14170952 PA14170954 PA14170956 PA14170961 PA14170962 PA14170963 PA14170964 PA14170965 PA14170966 PA14170967 PA14170968 PA14170969 PA14170970 PA14170971 PA14170973 PA14170974 PA14170975 PA14170976 PA14170977 PA14170978 PA14170979 PA14170980 PA14170981 PA14170982 PA14170983 PA14170984 PA14170985 PA14170986 PA14170987 PA14170988 PA14170989 PA14170990 PA14170991 PA14170992 PA14170993 PA14170994 PA14170995 PA14170996 PA14170997 PA14170998 PA14170999 PA14171000 PA15171001 PA15171002 PA15171003 PA15171004 PA15171005 PA15171006 PA15171007 PA15171008 PA15171009 PA15171010 PA15171011 PA15171012 PA15171013 PA15171014 PA15171015 PA15171016 PA15171017 PA15171018 PA15171019 PA15171020 PA15171021 PA15171022 PA15171023 PA15171024 PA15171025 PA15171026 PA15171027 PA15171028 PA15171029 PA15171030 PA15171031 PA15171032 PA15171033 PA15171034 PA15171035 PA15171036 PA15171037 PA15171038 PA15171039 PA15171040 PA15171041 PA15171042 PA15171043 PA15171044 PA15171045 PA15171046 PA15171047 PA15171048 PA22172187 PA23172256 PA23172257 PA23172258 PA23172259 PA23172260 PA23172261 PA23172262 PA23172263 PA23172264 PA23172265 PA23172266 PA23172267…

Reason for recall

The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory default settings when the monitor is powered off.

FDA-determined cause: Nonconforming Material/Component

Action taken

Urgent Medical Device Correction notification letters dated 7/31/18 were distributed to customers. The letters included the following instructions for customers: Share this notification with all care environments where the Capnostream" 20 and Capnostream" 20p bedside patient monitors are used, particularly in the areas where nurse call may be enabled (general care floors). If your facility has distributed these bedside patient monitors to other persons or facilities, please promptly forward a copy of this letter to those recipients. Medtronic is recommending continued use of Capnostream" devices. Users should confirm that the date/time is accurately displayed. If the date/time requires reset, all user defined settings, such as alarms and nurse call, will also require reset until the software is updated. Please follow the instruction in the Operators Manual to set user-defined default settings. The Operators Manual is available at http://www.medtronic.com/covidien/enus/support/product-manuals.html, refer to the Institutional Settings section. Return the completed Acknowledgement and Receipt Form by fax or email even if you have no inventory. In October 2018, download the software update from the Medtronic website via the following link which includes directions on how to download and install the software: http://www.medtronic.com/covidien/en-us/support/patient-monitoring-equipmentsoftware-upgrades.html.

Details

Recall number
Z-0043-2019
Initiated
July 31, 2018
Terminated
September 20, 2022
Product code
DQA – Oximeter
Quantity
192
Distribution
The products were distributed US nationwide and Puerto Rico. The products were distributed to the following foreign countries: Argentina, Australia, Austria, Bahrain, Belgium, Canada, Columbia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Kazakhstan, Kuwait, Mexico, Netherlands, Norway, Oman, Pakistan, Panama, Peru…
Recalling firm
Oridion Medical 1987 Ltd., Jerusalem, Har Hotzvi

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0020-2019 Class IICAPNOSTREAM 20 (US) N, Part Number CS08651The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0021-2019 Class IICAPNOSTREAM 20P (US) N, Part Number CS08651-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0022-2019 Class IICAPNOSTREAM 20P (US) N - ROHS, Part Number CS08651-02The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0023-2019 Class IICAPNOSTREAM 20P (US) N - ROHS, Part Number CS08651-03The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0024-2019 Class IICAPNOSTREAM 20 (US) N W/PRINTER, Part Number CS08657The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0025-2019 Class IICAPNOSTREAM 20P (US) N W/PRINTER - ROHS, Part Number CS08657-02The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0026-2019 Class IICAPNOSTREAM 20P (US) N W/PRINTER, Part Number CS08657-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0027-2019 Class IICAPNOSTREAM 20P (US) N W/PRINTER - ROHS, Part Number CS08657-03The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0028-2019 Class IICapnostream20 (US) M, Part Number CS08653The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0029-2019 Class IICapnostream20 (US) M W/PRINTER, Part Number CS08659The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0030-2019 Class IICapnostream20 (US) M TO MASIMO, Part Number CS08750The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0031-2019 Class IICapnostream20 (US) M W/PRINTER TO MASIMO, Part Number CS08751The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0032-2019 Class IICapnostream20 (US) N REFURBISHED, Part Number CS78651The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0033-2019 Class IICapnostream20 (US)M REFURBISHED, Part Number CS78653The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0034-2019 Class IICapnostream20 (US) N W/PRINTER REFURBISHED, Part Number CS78657The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0035-2019 Class IICapnostream20 (US) M W/PRINTER REFURBISHED, Part Number CS78659The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0036-2019 Class IICAPNOSTREAM 20 (INTL) N, Part Number CS08652The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0037-2019 Class IICAPNOSTREAM 20P (INTL) N, Part Number CS08652-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0038-2019 Class IICAPNOSTREAM 20P (INTL) N - ROHS, Part Number CS08652-02The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0039-2019 Class IICAPNOSTREAM 20P (INTL) N - ROHS, Part Number CS08652-03The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0040-2019 Class IICAPNOSTREAM 20 (INTL) N W/PRINTER, Part Number CS08658The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0041-2019 Class IICAPNOSTREAM 20P (INTL) N W/PRINTER, Part Number CS08658-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0042-2019 Class IICAPNOSTREAM 20P (INTL) N W/PRINTER - ROHS, CS08658-02The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0044-2019 Class IICapnostream20 (INTL) M, CS08654The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0045-2019 Class IICapnostream20 M ADULT ICU (CPM), CS08654RAThe date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0046-2019 Class IICapnostream20 M NICU (CPM), CS08654RNThe date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0047-2019 Class IICapnostream20 (INTL) M W/PRINTER, CS08660The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0048-2019 Class IICapnostream20 M W/PRINTER CYVITAL, CS08796The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0049-2019 Class IICapnostream20 N W/PRINTER CYVITAL, CS08798The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0050-2019 Class IICapnostream20 N CYVITAL, CS08799The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated