Recall Z-0028-2019

Oridion Medical 1987 Ltd. · initiated July 31, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Capnostream20 (US) M, Part Number CS08653

Codes / lots: B300001095 B300001167 B300001179 B300001180 B300001201 B300001203 B300001215 B300001216 B300001218 B300001195 B300001228 B300001364 B300001448 B300001407 B500001573 B500001539 B500001547 B500001548 B500001550 B500001553 B500001559 B500001563 B500001567 B500001571 B500001580 B500001581 B500001582 B500001583 B500001588 B500001592 B500001502 B500001503 B500001506 B500001515 B500001519 B500001523 B500001525 B500001528 B500001529 B500001533 B500001538 B300001446 B500001478 B500001507 B500001518 B500001530 B500001561 B500001568 B500001569 B500001587 B300001360 B300001373 B300001441 B300001500 B500001501 B500001450 B500001461 B500001479 B500001481 B500001484 B500001485 B500001486 B500001487 B500001491 B500001493 B500001494 B500001495 B500001496 B500001497 B500001542 B500001543 B300001268 B500001362 B500001383 B500001399 B500001482 B500001489 B500001490 B500001492 B500001498 B500001537 B500001544 B500001504 B500001513 B500001514 B300001405 B500001540 B500001541 B500001650 B500001651 B500001677 B300001409 B500001663 B500001684 B500001685 B500001602 B500001648 B500001656 B500001661 B500001662 B500001667 B500001669 B500001671 B500001672 B500001673 B500001676 B500001678 B500001679 B500001681…

Reason for recall

The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory default settings when the monitor is powered off.

FDA-determined cause: Nonconforming Material/Component

Action taken

Urgent Medical Device Correction notification letters dated 7/31/18 were distributed to customers. The letters included the following instructions for customers: Share this notification with all care environments where the Capnostream" 20 and Capnostream" 20p bedside patient monitors are used, particularly in the areas where nurse call may be enabled (general care floors). If your facility has distributed these bedside patient monitors to other persons or facilities, please promptly forward a copy of this letter to those recipients. Medtronic is recommending continued use of Capnostream" devices. Users should confirm that the date/time is accurately displayed. If the date/time requires reset, all user defined settings, such as alarms and nurse call, will also require reset until the software is updated. Please follow the instruction in the Operators Manual to set user-defined default settings. The Operators Manual is available at http://www.medtronic.com/covidien/enus/support/product-manuals.html, refer to the Institutional Settings section. Return the completed Acknowledgement and Receipt Form by fax or email even if you have no inventory. In October 2018, download the software update from the Medtronic website via the following link which includes directions on how to download and install the software: http://www.medtronic.com/covidien/en-us/support/patient-monitoring-equipmentsoftware-upgrades.html.

Details

Recall number
Z-0028-2019
Initiated
July 31, 2018
Terminated
September 20, 2022
Product code
DQA – Oximeter
Quantity
769
Distribution
The products were distributed US nationwide and Puerto Rico. The products were distributed to the following foreign countries: Argentina, Australia, Austria, Bahrain, Belgium, Canada, Columbia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Kazakhstan, Kuwait, Mexico, Netherlands, Norway, Oman, Pakistan, Panama, Peru…
Recalling firm
Oridion Medical 1987 Ltd., Jerusalem, Har Hotzvi

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0020-2019 Class IICAPNOSTREAM 20 (US) N, Part Number CS08651The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0021-2019 Class IICAPNOSTREAM 20P (US) N, Part Number CS08651-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0022-2019 Class IICAPNOSTREAM 20P (US) N - ROHS, Part Number CS08651-02The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0023-2019 Class IICAPNOSTREAM 20P (US) N - ROHS, Part Number CS08651-03The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0024-2019 Class IICAPNOSTREAM 20 (US) N W/PRINTER, Part Number CS08657The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0025-2019 Class IICAPNOSTREAM 20P (US) N W/PRINTER - ROHS, Part Number CS08657-02The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0026-2019 Class IICAPNOSTREAM 20P (US) N W/PRINTER, Part Number CS08657-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0027-2019 Class IICAPNOSTREAM 20P (US) N W/PRINTER - ROHS, Part Number CS08657-03The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0029-2019 Class IICapnostream20 (US) M W/PRINTER, Part Number CS08659The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0030-2019 Class IICapnostream20 (US) M TO MASIMO, Part Number CS08750The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0031-2019 Class IICapnostream20 (US) M W/PRINTER TO MASIMO, Part Number CS08751The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0032-2019 Class IICapnostream20 (US) N REFURBISHED, Part Number CS78651The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0033-2019 Class IICapnostream20 (US)M REFURBISHED, Part Number CS78653The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0034-2019 Class IICapnostream20 (US) N W/PRINTER REFURBISHED, Part Number CS78657The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0035-2019 Class IICapnostream20 (US) M W/PRINTER REFURBISHED, Part Number CS78659The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0036-2019 Class IICAPNOSTREAM 20 (INTL) N, Part Number CS08652The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0037-2019 Class IICAPNOSTREAM 20P (INTL) N, Part Number CS08652-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0038-2019 Class IICAPNOSTREAM 20P (INTL) N - ROHS, Part Number CS08652-02The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0039-2019 Class IICAPNOSTREAM 20P (INTL) N - ROHS, Part Number CS08652-03The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0040-2019 Class IICAPNOSTREAM 20 (INTL) N W/PRINTER, Part Number CS08658The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0041-2019 Class IICAPNOSTREAM 20P (INTL) N W/PRINTER, Part Number CS08658-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0042-2019 Class IICAPNOSTREAM 20P (INTL) N W/PRINTER - ROHS, CS08658-02The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0043-2019 Class IICAPNOSTREAM 20P (INTL) N W/PRINTER - ROHS, CS08658-03The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0044-2019 Class IICapnostream20 (INTL) M, CS08654The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0045-2019 Class IICapnostream20 M ADULT ICU (CPM), CS08654RAThe date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0046-2019 Class IICapnostream20 M NICU (CPM), CS08654RNThe date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0047-2019 Class IICapnostream20 (INTL) M W/PRINTER, CS08660The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0048-2019 Class IICapnostream20 M W/PRINTER CYVITAL, CS08796The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0049-2019 Class IICapnostream20 N W/PRINTER CYVITAL, CS08798The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0050-2019 Class IICapnostream20 N CYVITAL, CS08799The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated