Recall Z-0022-2019

Oridion Medical 1987 Ltd. · initiated July 31, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

CAPNOSTREAM 20P (US) N - ROHS, Part Number CS08651-02

Codes / lots: B836141101 B836141102 B836141103 B836141104 B836141106 B836141107 B836141108 B836141109 B836141110 B836141111 B836141113 B836141116 B836141117 B836141118 B836141119 B836141120 B836141121 B836141122 B836141123 B836141124 B836141125 B836141126 B836141127 B836141128 B836141130 B836141131 B836141133 B836141134 B836141135 B836141136 B836141137 B836141138 B836141139 B836141140 B836141141 B836141142 B836141143 B836141144 B836141145 B836141146 B836141147 B836141148 B836141149 B836141150 B836141151 B836141152 B836141153 B836141154 B836141155 B836141156 B836141157 B836141158 B836141159 B836141160 B836141161 B836141162 B836141163 B836141164 B836141165 B836141166 B836141167 B836141169 B836141170 B836141172 B836141173 B836141174 B836141175 B836141176 B836141177 B836141105 B836141112 B836141179 B836141180 B836141181 B836141182 B836141184 B836141185 B836141186 B836141187 B836141188 B836141189 B836141190 B836141192 B836141193 B836141195 B836141197 B836141199 B836141200 B836141201 B836141203 B836141204 B836141205 B836141208 B836141211 B836141213 B836141215 B836141216 B836141219 B836141220 B836141227 B836141228 B837141233 B837141234 B837141235 B837141237 B837141240 B837141241 B837141242 B837141246…

Reason for recall

The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory default settings when the monitor is powered off.

FDA-determined cause: Nonconforming Material/Component

Action taken

Urgent Medical Device Correction notification letters dated 7/31/18 were distributed to customers. The letters included the following instructions for customers: Share this notification with all care environments where the Capnostream" 20 and Capnostream" 20p bedside patient monitors are used, particularly in the areas where nurse call may be enabled (general care floors). If your facility has distributed these bedside patient monitors to other persons or facilities, please promptly forward a copy of this letter to those recipients. Medtronic is recommending continued use of Capnostream" devices. Users should confirm that the date/time is accurately displayed. If the date/time requires reset, all user defined settings, such as alarms and nurse call, will also require reset until the software is updated. Please follow the instruction in the Operators Manual to set user-defined default settings. The Operators Manual is available at http://www.medtronic.com/covidien/enus/support/product-manuals.html, refer to the Institutional Settings section. Return the completed Acknowledgement and Receipt Form by fax or email even if you have no inventory. In October 2018, download the software update from the Medtronic website via the following link which includes directions on how to download and install the software: http://www.medtronic.com/covidien/en-us/support/patient-monitoring-equipmentsoftware-upgrades.html.

Details

Recall number
Z-0022-2019
Initiated
July 31, 2018
Terminated
September 20, 2022
Product code
DQA – Oximeter
Quantity
10356
Distribution
The products were distributed US nationwide and Puerto Rico. The products were distributed to the following foreign countries: Argentina, Australia, Austria, Bahrain, Belgium, Canada, Columbia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Kazakhstan, Kuwait, Mexico, Netherlands, Norway, Oman, Pakistan, Panama, Peru…
Recalling firm
Oridion Medical 1987 Ltd., Jerusalem, Har Hotzvi

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-0020-2019 Class IICAPNOSTREAM 20 (US) N, Part Number CS08651The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0021-2019 Class IICAPNOSTREAM 20P (US) N, Part Number CS08651-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0023-2019 Class IICAPNOSTREAM 20P (US) N - ROHS, Part Number CS08651-03The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0024-2019 Class IICAPNOSTREAM 20 (US) N W/PRINTER, Part Number CS08657The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0025-2019 Class IICAPNOSTREAM 20P (US) N W/PRINTER - ROHS, Part Number CS08657-02The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0026-2019 Class IICAPNOSTREAM 20P (US) N W/PRINTER, Part Number CS08657-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0027-2019 Class IICAPNOSTREAM 20P (US) N W/PRINTER - ROHS, Part Number CS08657-03The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0028-2019 Class IICapnostream20 (US) M, Part Number CS08653The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0029-2019 Class IICapnostream20 (US) M W/PRINTER, Part Number CS08659The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0030-2019 Class IICapnostream20 (US) M TO MASIMO, Part Number CS08750The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0031-2019 Class IICapnostream20 (US) M W/PRINTER TO MASIMO, Part Number CS08751The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0032-2019 Class IICapnostream20 (US) N REFURBISHED, Part Number CS78651The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0033-2019 Class IICapnostream20 (US)M REFURBISHED, Part Number CS78653The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0034-2019 Class IICapnostream20 (US) N W/PRINTER REFURBISHED, Part Number CS78657The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0035-2019 Class IICapnostream20 (US) M W/PRINTER REFURBISHED, Part Number CS78659The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0036-2019 Class IICAPNOSTREAM 20 (INTL) N, Part Number CS08652The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0037-2019 Class IICAPNOSTREAM 20P (INTL) N, Part Number CS08652-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0038-2019 Class IICAPNOSTREAM 20P (INTL) N - ROHS, Part Number CS08652-02The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0039-2019 Class IICAPNOSTREAM 20P (INTL) N - ROHS, Part Number CS08652-03The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0040-2019 Class IICAPNOSTREAM 20 (INTL) N W/PRINTER, Part Number CS08658The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0041-2019 Class IICAPNOSTREAM 20P (INTL) N W/PRINTER, Part Number CS08658-01The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0042-2019 Class IICAPNOSTREAM 20P (INTL) N W/PRINTER - ROHS, CS08658-02The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0043-2019 Class IICAPNOSTREAM 20P (INTL) N W/PRINTER - ROHS, CS08658-03The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0044-2019 Class IICapnostream20 (INTL) M, CS08654The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0045-2019 Class IICapnostream20 M ADULT ICU (CPM), CS08654RAThe date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0046-2019 Class IICapnostream20 M NICU (CPM), CS08654RNThe date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0047-2019 Class IICapnostream20 (INTL) M W/PRINTER, CS08660The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0048-2019 Class IICapnostream20 M W/PRINTER CYVITAL, CS08796The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0049-2019 Class IICapnostream20 N W/PRINTER CYVITAL, CS08798The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated
Z-0050-2019 Class IICapnostream20 N CYVITAL, CS08799The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory… 07/31/2018Terminated